Specifications
Table 21: Spurious response with signal at center frequency (offset
≥
400 kHz)
Span
≤
25 MHz,
Swept Spans >
25 MHz
For Option B40
1
Swept Spans >
40 MHz
For Option B85, B125, B16x
1
40 MHz < Span
≤
160 MHz
For Option
B85HD,
B125HD,
B16xHD
1
40 MHz < Span
≤
165 MHz
Frequency
Speci
fi
cationTypical
Speci
fi
cationTypical
Speci
fi
cation
Typical
Typical
1 MHz to 32 MHz (LF
band)
–80 dBc
–85 dBc
– –
– –
– –
– –
– –
30 MHz to 3 GHz
–73 dBc
–80 dBc
–73 dBc
–80 dBc
–73 dBc
–75 dBc
–80 dBc
3 GHz to 6.2 GHz
(RSA5106B)
–73 dBc
–80 dBc
–73 dBc
–80 dBc
–73 dBc
–75 dBc
–80 dBc
6.2 GHz to 15 GHz
(RSA5115B)
–70 dBc
–80 dBc
–70 dBc
–80 dBc
–70 dBc
–73 dBc
–80 dBc
15 GHz to 26 GHz
(RSA5126B)
–66 dBc
–76 dBc
–66 dBc
–76 dBc
–66 dBc
–73 dBc
–76 dBc
1
Center frequency
≥
150 MHz for Options B40/B85/B125/B16x.
Table 22: Spurious response with signal at center frequency (10 kHz
≤
offset
≤
400 kHz)
Frequency
Typical, (Span = 1 MHz)
1 MHz to 32 MHz (LF band)
–75 dBc
30 MHz to 3 GHz
–75 dBc
3 GHz to 6.2 GHz (RSA5126B only)
–75 dBc
15 GHz to 26.5 GHz (RSA5126B only)
–68 dBc
Table 23: Spurious response with signal at input other than center frequency
Span
≤
25 MHz,
swept spans
>25 MHz
For Option
B40
1
Span
≤
40 MHz,
swept spans
>40 MHz
For Option
B85
1
40 MHz
< Span
≤
85 MHz
For Option
B125, B16x
1 2
85 MHz < Span
≤
165 MHz
For Option
B85HD,
B125HD,
B16xHD
1
40 MHz < Span
≤
165 MHz
Frequency
Typical
Typical
Typical
Typical
Typical
1 MHz to 32 MHz (LF band)
–80 dBc
– –
– –
– –
– –
30 MHz to 3 GHz
–80 dBc
–80 dBc
–76 dBc
–73 dBc
–80 dBc
3 GHz to 6.2 GHz (RSA5106B)
–80 dBc
–80 dBc
–76 dBc
–73 dBc
–80 dBc
6.2 GHz to 15 GHz (RSA5115B)
–80 dBc
–80 dBc
–73 dBc
–73 dBc
–80 dBc
15 GHz to 26.5 GHz (RSA5126B)
–76 dBc
–76 dBc
–73 dBc
–73 dBc
–76 dBc
1
Center frequency
≥
150 MHz for Options B40/B85/B125/B16x.
2
-70 dBc for input signals 20 MHz above or below instrument center frequency.
3
Mean of performance measured over a sample interval.
RSA5100B Series Technical Reference