Performance Verification
Test Record
Print out the following test record pages and use them to record the performance
test results for your signal analyzer.
RSA5100B Series Test
Record
Model:
Certi
fi
cate Number:
Technician:
Serial Number:
Calibration Date:
Frequency Accuracy
Low limit
Test Result
High limit
Reference output frequency accuracy
(Standard)
9,999,960 Hz
10,000,040 Hz
(Option PFR)
9,999,996 Hz
10,000,004 Hz
Reference output power level
0 dBm
External reference input level
Fail
Pass
Phase Noise (with Option 11)
Test Result
High limit
Offset
CF + 1 kHz
–103 dBc/Hz
CF + 10 kHz
–109 dBc/Hz
CF + 100 kHz
–112 dBc/Hz
CF + 1 MHz
–130 dBc/Hz
CF + 6 MHz
–137 dBc/Hz
CF + 10 MHz
–137 dBc/Hz
Phase Noise (without Option 11)
Cursor
Carrier power
Phase noise
1
High limit
Offset
CF + 1 kHz
–103 dBc/Hz
CF + 10 kHz
–109 dBc/Hz
CF + 100 kHz
–112 dBc/Hz
CF + 1 MHz
–130 dBc/Hz
CF + 6 MHz
–137 dBc/Hz
CF + 10 MHz
–137 dBc/Hz
1
Phase noise = –(cursor measurement – carrier power)
134
RSA5100B Series Technical Reference