Specifications
Table 17: Pulse measurements, typical (cont.)
Characteristic
Description
Chirp Frequency Linearity
(Absolute Frequency Error RMS)
60 MHz BW
13
20 MHz BW
14
2 GHz
±25 kHz
±3 kHz
10 GHz
±30 kHz
±3 kHz
20 GHz
±30 kHz
±5 kHz
85 MHz BW
13
25 and 40 MHz BW
14
2 GHz
±25 kHz
±5 kHz
10 GHz
±30 kHz
±8 kHz
20 GHz
±30 kHz
±10 kHz
125, 160 MHz BW
13
2 GHz
±35 kHz
10 GHz
±40 kHz
20 GHz
±40 kHz
1
For conditions of:
CW (non-chirped) pulses
Frequency Estimation = Manual
Pulse ON power
≥
–20 dBm
Signal peak at Ref Lvl.
Atten = 0 dB
Pulse width
≥
200 ns.
PRI
≤
300 us.
Duty cycle
≥
0.0007
t
meas
– t
reference
≤
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured from the
50% point of the Tr or Tf.
2
For conditions of:
CW (non-chirped) pulses
Frequency Estimation = Manual
Pulse ON power
≥
–20 dBm
Signal peak at Ref Lvl.
Atten = 0 dB
Pulse width
≥
300 ns.
PRI
≤
300 us.
Duty cycle
≥
0.001
t
meas
– t
reference
≤
10 ms
Measurement time position excludes the beginning and ending of the pulse extending for a time = (10/measurement Bandwidth) as measured from the
50% point of the Tr or Tf.
3
CW (non-chirped) pulses
Frequency Estimation = Manual
Pulse ON power
≥
–20 dBm
Signal peak at Ref Lvl.
Atten = 0 dB
Pulse width
≥
1
μ
s.
PRI
≤
300 us.
RSA5100B Series Technical Reference
21