Specifications
Performance Conditions
The performance limits in these specifications are valid with these conditions:
The signal analyzer must have been calibrated and adjusted at an ambient
temperature b18 °C and +28 °C.
The signal analyzer must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications.
Ambient temperature must remain stable while executing the self-alignment.
The signal analyzer must have had a warm-up period of at least 20 minutes
after starting the RSA5100B application.
Electrical Speci
fi
cations
Table 2: Frequency
Characteristic
Description
Measurement frequency
LF band
1 Hz to 32 MHz
RSA5103B (RF band)
9 kHz to 3 GHz
RSA5106B (RF band)
9 kHz to 6.2 GHz
RSA5115B (RF band)
1 MHz to 15 GHz
Frequency
range, nominal
RSA5126B (RF band)
1 MHz to 26.5 GHz
Readout Accuracy
±(RE × MF + 0.001 × Span + 2) Hz
RE: Reference Frequency Error
MF: Marker Frequency [Hz]
Frequency
Marker
Readout Resolution
Reference level dependent
As small as 0.0001 µV
Residual FM, typical
<2 Hz
p-p
in 1 second at 200 MHz CF, 100 Hz span, Freq vs Time
mode, Autoscale (95% con
fi
dence)
Span Accuracy
±0.3% of span (Auto mode)
2
RSA5100B Series Technical Reference