Speci
fi
cations
Table 1: General characteristics (cont.)
Characteristic
Description
Maximum Working Voltage
15 V (DC + peak AC)
Spurious Input Signals
To meet modulation measurement speci
fi
cations:
(Total power of interest - Total power in spurious) > 35 dB
or
| Center frequency of signal of interest - center frequency of spurious signal | > 1.4 MHz
and
(Total power of interest - Total power in all spurious)
≥
–3 dB
External Frequency
Reference Input
Impedance
1500 ohms (nominal)
Input Connector
50 ohm BNC
Frequency Range
Any multiple of 1 MHz up to 15 MHz inclusive, ± 1 ppm
Any multiple of 1.2288 MHz up to 19.6608 MHz inclusive, ± 1 ppm
2.048 MHz ± 1 ppm
4.8 MHz ± 1 ppm
Level Range
–15 dBm to +15 dBm
dBm levels assume 50 ohm source
72