Troubleshooting
P6249 4 GHz 5X Active Probe Instruction Manual
55
Table 9: Troubleshooting guide (Cont.)
Symptom
Action
Result
Checks
No Offset Adjust
(Signal present)
Go to
Probe Power
row
Probe power OK
Continue checks
Check variation of
offset input from host
instrument. See
Figure 21. Signal
should vary between
+1 V and --1 V.
Offset input OK
Offset input bad
Replace probe.
Check TEKPROBE
interface pins on the
compensation box.
See page 47.
Check TEKPROBE
interface on host
instrument. (Try an
adjacent channel.)
Repair host
instrument
Probe Power
Check the power
test points in the
compensation box.
See Figure 21.
Power OK
Continue checks.
Power bad
Check TEKPROBE
interface pins on the
compensation box.
See page 47.
Check TEKPROBE
interface on host
instrument. (Try an
adjacent channel.)
Repair host
instrument
Summary of Contents for P6249
Page 1: ...Instruction Manual P6249 4 GHz 5X Active Probe 071 0538 05 www tektronix com...
Page 4: ......
Page 8: ...Table of Contents iv P6249 4 GHz 5X Active Probe Instruction Manual...
Page 28: ...Functional Check 16 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 34: ...Helpful Hints 22 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 40: ......
Page 44: ...Theory of Operation 32 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 54: ...Performance Verification 42 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 64: ...Maintenance 52 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 68: ...Troubleshooting 56 P6249 4 GHz 5X Active Probe Instruction Manual...