P6249 4 GHz 5X Active Probe Instruction Manual
23
Specifications
These specifications apply to a P6249 probe when used with a
TDS 694C oscilloscope.
The probe and oscilloscope must first be allowed to warm up for
20 minutes before measurements are taken.
CAUTION.
Do not apply voltages beyond the non-destructive input
voltage range to the probe. Damage to the probe or circuit under test
may result.
Table 4: Warranted electrical specifications
DC Attenuation Accuracy (probe only)
0.2
±
2% (excludes offset error)
Output Zero
±
10 mV or less at output of probe
Rise Time (probe only)
≤
120 ps
Table 5: Typical electrical characteristics
Bandwidth, (probe only)
4 GHz (See Figure 8)
Linear Input Dynamic Range
--2.0 V to +2.0 V.
(Equivalent to --0.4 V to +0.4 V
at the output of the probe.)
Linearity
ᐔ
0.1% over a dynamic range of --1.75 V to
+1.75 V
ᐔ
1.0% over a dynamic range of --2.0 V to
+2.0 V
Nondestructive Input Voltage Range
--30 V to +30 V (DC + peak AC)
Input Resistance
20 k
Ω
at DC. (See Figure 7)
Input Capacitance
1.0 pF
Summary of Contents for P6249
Page 1: ...Instruction Manual P6249 4 GHz 5X Active Probe 071 0538 05 www tektronix com...
Page 4: ......
Page 8: ...Table of Contents iv P6249 4 GHz 5X Active Probe Instruction Manual...
Page 28: ...Functional Check 16 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 34: ...Helpful Hints 22 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 40: ......
Page 44: ...Theory of Operation 32 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 54: ...Performance Verification 42 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 64: ...Maintenance 52 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 68: ...Troubleshooting 56 P6249 4 GHz 5X Active Probe Instruction Manual...