Performance Verification
40
P6249 4 GHz 5X Active Probe Instruction Manual
11.
Adjust the oscilloscope horizontal and vertical position controls
to display a signal similar to that shown in Figure 13.
NOTE
. Do not touch the probe tip adapter when making calibration
measurements. Measurement accuracy is degraded when the probe
tip adapter is handled.
12.
Use the oscilloscope measurement capability to display rise time.
Rise time is determined from the 10% and 90% amplitude points
on the waveform. Record the rise time as
t
s+p
.
13.
Calculate the probe only rise time using the following formula:
t
p
=
t
(
s
+
p
)
2
−
t
s
2
Ꭹ
14.
Check that the calculated rise time meets the probe specification.
The rise time (
t
p
) must be
≤
120 ps.
15.
Record the results on the test record.
Summary of Contents for P6249
Page 1: ...Instruction Manual P6249 4 GHz 5X Active Probe 071 0538 05 www tektronix com...
Page 4: ......
Page 8: ...Table of Contents iv P6249 4 GHz 5X Active Probe Instruction Manual...
Page 28: ...Functional Check 16 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 34: ...Helpful Hints 22 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 40: ......
Page 44: ...Theory of Operation 32 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 54: ...Performance Verification 42 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 64: ...Maintenance 52 P6249 4 GHz 5X Active Probe Instruction Manual...
Page 68: ...Troubleshooting 56 P6249 4 GHz 5X Active Probe Instruction Manual...