Adjustments
46
P6209 4 GHz Probe Instruction Manual
Offset Scale
Use this procedure to adjust the offset scale of the probe. The offset
zero of the probe should be checked before making any adjustment
to the offset scale. Refer to Figure 17 on page 44 for the location of
the test points and adjustments.
1.
Connect the probe power cable to the oscilloscope.
2.
Set the oscilloscope trigger source to internal clock.
3.
Set the oscilloscope channel vertical offset to +2.5 V.
4.
Ground the probe tip by connecting the two KlipChip leads
together.
5.
Using the (--) lead of the DMM, probe the ground lug in the probe
compensation box.
6.
Using the (+) lead of the DMM, carefully probe pin 7 of J2 (white
wire) in the probe compensation box.
NOTE
. The offset voltage measured on pin 7 will be of opposite
polarity and approximately one fifth of the offset value displayed on
the oscilloscope.
7.
Adjust the oscilloscope offset so that pin 7 of J2 measures
--0.50 V on the DMM.
8.
Set the DC power supply to +2.5 V, as measured on the DMM.
9.
Connect the equipment as shown in Figure 19. Take care not to
disturb the offset level setting.
10.
Adjust the
Offset Scale
(R3) in the probe compensation box so
the output measures 0.000 V,
±
1 mV on the DMM.
Summary of Contents for P6209
Page 5: ......
Page 9: ...Table of Contents iv P6209 4 GHz Probe Instruction Manual...
Page 21: ...Product Description 8 P6209 4 GHz Probe Instruction Manual...
Page 29: ...Operating Basics 16 P6209 4 GHz Probe Instruction Manual...
Page 33: ...Helpful Hints 20 P6209 4 GHz Probe Instruction Manual...
Page 41: ......
Page 45: ...Theory of Operation 32 P6209 4 GHz Probe Instruction Manual...
Page 55: ...Performance Verification 42 P6209 4 GHz Probe Instruction Manual...
Page 61: ...Adjustments 48 P6209 4 GHz Probe Instruction Manual...
Page 63: ...Maintenance 50 P6209 4 GHz Probe Instruction Manual...
Page 67: ...Troubleshooting 54 P6209 4 GHz Probe Instruction Manual...
Page 75: ...Replaceable Parts P6209 4 GHz 5X Active Probe 62...