Troubleshooting
P6209 4 GHz Probe Instruction Manual
53
Table 6: Troubleshooting guide
Symptom
Checks
Result
Action
No signal out
Check known signal
Go to
No Probe
Power
No signal.
Probe power OK.
Continue checks
Replace probe.
Distorted or
nonlinear signal
Is the signal within
linear dynamic
range? See page 13.
No
Set channel vertical
offset to match signal
characteristic.
No Offset Adjust
(Signal present)
Go to
Probe Power
row
Probe power OK
Continue checks
Check variation of
offset input from host
instrument. See
Figure 20. Offset
control voltage at pin
7 of J2 should vary
b1 V and
--1 V.
Offset input OK
Replace probe.
No Probe Power
Check the power
test points in the
compensation box.
See Figure 20.
Power OK
Continue checks.
Power bad
Check for continuity
between the
TEKPROBE SMA
connector pins and
the test points in the
compensation box.
See Figures 20 and
21 on page 52.
Try an adjacent
channel.
Repair host
instrument
Summary of Contents for P6209
Page 5: ......
Page 9: ...Table of Contents iv P6209 4 GHz Probe Instruction Manual...
Page 21: ...Product Description 8 P6209 4 GHz Probe Instruction Manual...
Page 29: ...Operating Basics 16 P6209 4 GHz Probe Instruction Manual...
Page 33: ...Helpful Hints 20 P6209 4 GHz Probe Instruction Manual...
Page 41: ......
Page 45: ...Theory of Operation 32 P6209 4 GHz Probe Instruction Manual...
Page 55: ...Performance Verification 42 P6209 4 GHz Probe Instruction Manual...
Page 61: ...Adjustments 48 P6209 4 GHz Probe Instruction Manual...
Page 63: ...Maintenance 50 P6209 4 GHz Probe Instruction Manual...
Page 67: ...Troubleshooting 54 P6209 4 GHz Probe Instruction Manual...
Page 75: ...Replaceable Parts P6209 4 GHz 5X Active Probe 62...