Probing Principles
Ampere-Second Product is the maximum
width of pulsed current that you can
measure when the pulse amplitude is
between the maximum continuous and
maximum pulsed current speci
fi
cations.
The maximum continuous speci
fi
cation
varies by frequency.
To determine if your measurement exceeds
the Ampere-second product, you must
fi
rst
determine the maximum allowable pulse width
or maximum allowable pulse amplitude, as
described in the following section.
P6021A Current Probe Instruction Manual
18