Adjustments
Midband Accuracy
This procedure describes the midband accuracy adjustments for the probe. You must complete the
Midband Accuracy Test
in the
Performance Veri
fi
cation
procedure before making this adjustment. Keep the test setup intact to perform the adjustment.
1.
Set the calibrator output to 2 V
RMS
@ 50 kHz.
2.
Set the digital multimeter (DMM) to measure AC volts,
≥
2.0 V range.
3.
Set the probe to the 2 mA/mV range.
4.
Measure the output amplitude on the DMM.
5.
Adjust the 2 mA/mV control to display 20.0 mV,
±0.6 mV on the DMM.
6.
Set the probe to the 10 mA/mV range.
7.
Adjust the 10 mA/mV control to display 4.0 mV,
±0.12 mV on the DMM.
8.
Disconnect the probe from the calibrator and disable
the output.
53
P6021A Current Probe Instruction Manual