5.
Connect the probe to any channel (1–4).
6.
Set the probe to the 2 mA/mV range.
7.
Clamp the current probe around the HF
current loop. Verify that the arrow indicator
on the probe points away from the pulse
generator.
8.
The rise time is displayed on the
oscilloscope. Record the rise time
measurement in the Test Record.
9.
Set the probe to the 10 mA/mV range and
measure the rise time. Record the rise time
measurement in the Test Record.
Bandwidth High Limit
10.
Calculate the high frequency limit of the
probe for each range setting using the
measured rise times in the formula shown:
11.
Record the calculated bandwidth limit
values in the Test Record.