Performance Veri
fi
cation
Test
1.
Set the DMM to measure AC volts, on
≥
2 volt range (or use Autoset).
2.
Set the calibration generator output to
2 V
RMS
@ 50 kHz, sine wave.
3.
Measure the generator output on the DMM
and record the output as M1.
This voltage measurement is used to
calculate the actual test current,
Itest
.
4.
Disconnect the cable at the BNC adapter
attached to the DMM.
5.
Connect the probe output to the BNC
adapter attached to the DMM.
6.
Clamp the current probe around the
calibration
fi
xture as shown. The arrow
indicator on the probe should point away
from the (+) terminal of the current source.
7.
Set the probe to the 2 mA/mV range.
8.
Set the DMM to the 100 mV range and
record the exact measurement of the DMM
as M2.
41
P6021A Current Probe Instruction Manual