Performance Veri
fi
cation
Test
1.
Enable the generator output.
2.
Use the oscilloscope cursors (shown in the
fi
gure) or the Fall Time Measurement to
measure the time constant (
τ
), beginning
from the peak of the rising edge (100%),
down to the 37% amplitude level of the
output waveform.
3.
Record the time constant (
τ
) in the Test
Record.
4.
Calculate the low frequency limit of the
probe bandwidth, using the time constant
in the formula shown:
5.
Record the calculated low frequency limit
value for the 2 mA/mV range in the Test
Record.
P6021A Current Probe Instruction Manual
46