Adjustments
2–4
P6015A Instruction Manual
Table 2–1: Test Equipment Required for Long-Form Adjustment
Item
Minimum
Requirements
Recommended
Example
Application
Oscilloscope
Input Impedance:1 M
W
Sensitivity: 1 mV/div
Bandwidth:
≥
100 MHz
Tektronix 11402A
with 11A32 plug-
in, or Tektronix
TDS 460
All checks and
adjustments
Calibration
Generator
Rise time:
≤
10 ns
Repetition rate: 1 MHz
Amplitude:
≥
50 V
Tektronix
PG 506A
1,2
All checks and
adjustments other
than bandwidth
Leveled Sine
Wave Generator
Range: 50 kHz to 75 MHz
Amplitude:
≥
5 V
Tektronix SG 503
2
Bandwidth check
BNC Male-to-GR
Adapter
Tektronix
part number
017-0064-00
All checks and
adjustments
BNC 50
W
Termination
Tektronix
part number
011-0049-01
Transient
response check
and adjustment,
bandwidth check
1
Oscilloscopes with less than 1 mV/div sensitivity can be used while
making most checks and adjustments. However, because of the 1000X
probe attenuation, the system will not display sufficient deflection for
optimum adjustment of transient response unless a calibration generator
having higher amplitude output is substituted.
2
Requires TM 500 or TM 5000 Series Power Module or equivalent.
Summary of Contents for P6015A
Page 4: ......
Page 12: ...Service Safety Summary viii P6015A Instruction Manual...
Page 14: ...Service Safety Summary x P6015A Instruction Manual...
Page 15: ...User Information...
Page 16: ......
Page 18: ...Overview 1 2 P6015A Instruction Manual Guard Ring Figure 1 1 The P6015A High Voltage Probe...
Page 38: ...In Detail 1 22 P6015A Instruction Manual...
Page 46: ......
Page 47: ...Service Information...
Page 48: ......
Page 50: ...Performance Verification 2 2 P6015A Instruction Manual...
Page 73: ...Index...
Page 74: ......