In Detail
1–16
P6015A Instruction Manual
Table 1–2: Test Equipment Required for Short-Form Adjustment
Item
Minimum
Requirements
Recommended
Example
Application
Oscilloscope
Input Impedance:1 M
W
Sensitivity: 1 mV/div
Bandwidth:
≥
100 MHz
Tektronix 11402A
with 11A32 plug-
in, or Tektronix
TDS 460
All adjustments
Calibration
Generator
Rise time:
≤
10 ns
Repetition rate: 1 MHz
Amplitude:
≥
50 V
Tektronix
PG 506A
1,2
All adjustments
BNC Male-to-GR
Adapter
Tektronix
part number
017-0064-00
All adjustments
BNC 50
W
Termination
Tektronix
part number
011-0049-01
Transient
response
adjustment
1
Oscilloscopes with less than 1mV/div sensitivity can be used while making
most adjustments. However, because of the 1000X probe attenuation, the
system will not display sufficient deflection for optimum adjustment of
transient response unless a calibration generator having higher amplitude
output is substituted.
2
Requires TM 500 or TM 5000 Series Power Module or equivalent.
Summary of Contents for P6015A
Page 4: ......
Page 12: ...Service Safety Summary viii P6015A Instruction Manual...
Page 14: ...Service Safety Summary x P6015A Instruction Manual...
Page 15: ...User Information...
Page 16: ......
Page 18: ...Overview 1 2 P6015A Instruction Manual Guard Ring Figure 1 1 The P6015A High Voltage Probe...
Page 38: ...In Detail 1 22 P6015A Instruction Manual...
Page 46: ......
Page 47: ...Service Information...
Page 48: ......
Page 50: ...Performance Verification 2 2 P6015A Instruction Manual...
Page 73: ...Index...
Page 74: ......