Specifications
1–24
P6015A Instruction Manual
Table 1–3: Warranted Electrical Characteristics
Characteristic
Information
Maximum input voltage
DC + peak AC
1
1.5 kV to 20 kV. See frequency derating curve in
Figure 1–4. (DC plus peak AC rating is limited to
temperatures below 35
°
C.)
Peak pulse
40 kV
a
(Never exceed 20 kV rms)
Duty cycle derating – 100 ms maximum duration at 10%
maximum duty cycle. See duration and duty cycle
derating curve in Figure 1–5.
Altitude derating – Peak pulse derated linearly from
40 kV at 8000 feet (2440 m) to 30 kV at 15,000 feet
(4570 m) altitude.
Relative Humidity (RH) derating – Voltage derated with
increasing temperature and relative humidity (see
Figure 1–7).
Bandwidth (–3 dB)
Test conditions: Test oscilloscope bandwidth must be
≥
100 MHz, Z
source
= 25
W
10-ft cable
75 MHz
25-ft cable
25 MHz
Rise Time
2
10-ft cable
≤
4.67 ns (calculated from bandwidth)
25-ft cable
≤
14 ns (calculated from bandwidth)
DC attenuation
1000:1
±
3% (Excluding
oscilloscope error)
Test conditions: Oscillo-
scope input resistance
must be 1 M
W
±
2%
1
Characteristic not checked in manual
2
T
r
(ns) = .35/BW (MHz)
Summary of Contents for P6015A
Page 4: ......
Page 12: ...Service Safety Summary viii P6015A Instruction Manual...
Page 14: ...Service Safety Summary x P6015A Instruction Manual...
Page 15: ...User Information...
Page 16: ......
Page 18: ...Overview 1 2 P6015A Instruction Manual Guard Ring Figure 1 1 The P6015A High Voltage Probe...
Page 38: ...In Detail 1 22 P6015A Instruction Manual...
Page 46: ......
Page 47: ...Service Information...
Page 48: ......
Page 50: ...Performance Verification 2 2 P6015A Instruction Manual...
Page 73: ...Index...
Page 74: ......