![Tektronix Keithley 2380-500-30 Performance Verification Manual Download Page 7](http://html1.mh-extra.com/html/tektronix/keithley-2380-500-30/keithley-2380-500-30_performance-verification-manual_1077859007.webp)
2380-500-30-905-01 Rev. A / November 2015
1
This section contains specifications for the Series 2380 and Series 2380J Programmable DC
electronic load. All specifications are guaranteed unless noted as "typical." Typical specifications
are provided for your convenience but are not guaranteed. Specifications that are marked with the
symbol
√
are checked in performance verification.
Model 2380-500-30 / 2380J-500-30 specifications
Parameter
Description
Constant Voltage Operation, nominal
This value can be set as a constant voltage
Voltage Range
Low range: 0~50 V, High range:0~500 V
Voltage Resolution, nominal
1~10 mV.
√
Voltage Setting Accuracy
0.05% FS ± 0.025% at 25 ºC ± 5 ºC
Constant Current Operation, nominal
This value can be set as a constant current
Current Range
Low range: 0~3 A, High range: 0~30 A
Current Resolution, nominal
0.1~1 mA
√
Current Setting Accuracy
+0.05% FS ± 0.05% at 25 °C ± 5 °C
T1
&
T2
20 uS~3600 S /Res:1 uS
Accuracy of transient mode
5 uS ± 100 ppm
rise/fall slope
Low range: 0.0001~0.1 A/uS, High range: 0.001~1 A/uS
Minimum rising time
≒
10 uS
Constant Resistance Operation, nominal
This value can be set as a constant resistance
Resistance Range
Low range :0.15~10
Ω
, High range: 10
Ω
~7.5 K
Ω
Resistance Resolution, nominal
16 bit.
Resistance Setting Accuracy
Under the temperature of 25 ºC ± 5 ºC, the accuracy for low range is
0.01% + 0.08 S, the accuracy for high range is 0.01% + 0.0008 S
Constant Power Operation, nominal This value can be set as a constant power
Power Range
750 W
Power Resolution, nominal
10 mW
Power Setting Accuracy
0.2% FS ± 0.2% at 25 °C ± 5 °C
Meter, nominal
Internal readback capability: voltage, current and power are measured
internally
Voltage Readback Range
Low range: 0~50 V, High range: 0~500 V
Voltage Readback Resolution, nominal
1~10 mV
√
Voltage Readback Accuracy
0.025% FS ± 0.025% at 25 ºC ± 5 ºC
Current Readback Range
Low range: 0~3 A, High range:0~30 A
Current Readback Resolution, nominal
0.1~1 mA
√
Current Readback Accuracy
0.05% FS ± 0.05% at 25 °C ± 5 °C
Power Readback Range
750 W
Power Readback Resolution, nominal
10 mW
Power Readback Accuracy
0.2% FS ± 0.2% at 25 °C ± 5 °C
Section 2