Test
description
Expected
value
Measured
value
Low limit
High limit
Unit
Test
result
High current (22.5 A)
22.500
22.4995
22.4724
22.5249
A
Pass
High current (30 A)
30.000
29.9999
29.9683
30.0283
A
Pass
Set sense accuracy voltage (50 V range)
Low voltage (0.1 V)
0.100
0.099
0.075
0.125
V
Pass
Low voltage (12.5 V)
12.500
12.499
12.472
12.528
V
Pass
Low voltage (25 V)
25.000
24.999
24.969
25.031
V
Pass
Low voltage (37.5 V)
37.500
37.498
37.466
37.534
V
Pass
Low voltage (50 V)
50.000
49.999
49.963
50.038
V
Pass
Readback sense accuracy voltage (50 V range)
Low voltage (0.1 V)
0.100
0.099
0.087
0.112
V
Pass
Low voltage (12.5 V)
12.500
12.499
12.484
12.515
V
Pass
Low voltage (25 V)
25.000
24.998
24.980
25.017
V
Pass
Low voltage (37.5 V)
37.500
37.498
37.476
37.520
V
Pass
Low voltage (50 V)
50.000
49.999
49.974
50.024
V
Pass
Set sense accuracy voltage (500 V range)
High voltage (0.1 V)
0.10
0.09
-0.15
0.35
V
Pass
High voltage (125 V)
125.00
125.00
124.72
125.28
V
Pass
High voltage (250 V)
250.00
249.99
249.69
250.31
V
Pass
High voltage (375 V)
375.00
374.99
374.66
375.34
V
Pass
High voltage (500V)
500.00
500.01
499.63
500.38
V
Pass