![Tektronix Keithley 2380-500-30 Performance Verification Manual Download Page 16](http://html1.mh-extra.com/html/tektronix/keithley-2380-500-30/keithley-2380-500-30_performance-verification-manual_1077859016.webp)
Model 2380-500-30 Programmable Instruments
Section 3: Performance Verification
2380-500-30-905-01 Rev. A / November 2015
7
Test
description
Expected
value
Measured
value
Low limit
High limit
Unit
Test
result
Readback accuracy power (750 W range)
Power (0 W)
0.00
0.00
-1.50
1.50
W
Pass
Power (187.5 W)
187.50
187.44
185.55
189.30
W
Pass
Power (375 W)
375.00
374.79
372.52
377.02
W
Pass
Power (562.5 W)
562.50
562.36
559.71
564.96
W
Pass
Power (750 W)
750.00
749.57
746.55
752.54
W
Pass
Set accuracy resistance (10
Ω
range)
Low resistance (0.15
Ω
)
0.150
0.150
0.148
0.152
Ω
Pass
Low resistance (2.5
Ω
)
2.500
2.500
2.083
3.125
Ω
Pass
Low resistance (5
Ω
)
5.000
5.001
3.571
8.335
Ω
Pass
Low resistance (7.5
Ω
)
7.500
7.501
4.687
18.755
Ω
Pass
Low resistance (10
Ω
)
10.000
10.003
5.555
50.025
Ω
Pass
Set accuracy resistance (7500
Ω
range)
High resistance (10
Ω
)
10.000
10.003
9.920
10.082
Ω
Pass
High resistance (100
Ω
)
100.00
100.10
92.58
108.71
Ω
Pass
High resistance (1000
Ω
)
1000.0
1008.0
555.5
5002.5
Ω
Pass
High resistance (3000
Ω
)
3000.0
3120.4
882.3
------
Ω
Pass
High resistance (7500
Ω
)
7500.0
8146.8
1071.4
------
Ω
Pass