background image

Model 2380-500-30 Programmable Instruments 

Section 3: Performance Verification 

2380-500-30-905-01 Rev. A / November 2015 

Test

 

description

 

Expected

 

value

 

Measured

value

 

Low limit

 

High limit

 

Unit

 

Test

 

result

 

Readback sense accuracy voltage (500 V range)

 

High voltage (0.1 V) 

0.10 

0.09 

-0.03

0.22 

Pass 

High voltage (125 V) 

125.00 

124.99 

124.84 

125.15 

Pass 

High voltage (250 V) 

250.00 

249.98 

249.80 

250.18 

Pass 

High voltage (375 V) 

375.00 

374.99 

374.77 

375.20 

Pass 

High voltage (500V) 

500.00 

500.01 

499.76 

500.26 

Pass 

High voltage (500 V) 

500.00 

500.01 

499.76 

500.26 

Pass 

Set local accuracy voltage (50 V range)

 

Low voltage (0.1 V) 

0.100 

0.100 

0.075 

0.125 

Pass 

Low voltage (12.5 V) 

12.500 

12.500 

12.472 

12.528 

Pass 

Low voltage (25 V) 

25.000 

24.999 

24.969 

25.031 

Pass 

Low voltage (37.5 V) 

37.500 

37.499 

37.466 

37.534 

Pass 

Low voltage (50 V) 

50.000 

50.000 

49.963 

50.038 

Pass 

Readback local accuracy voltage (50 V range)

 

Low voltage (0.1 V) 

0.100 

0.099 

0.087 

0.112 

Pass 

Low voltage (12.5 V) 

12.500 

12.499 

12.484 

12.515 

Pass 

Low voltage (25 V) 

25.000 

24.999 

24.980 

25.018 

Pass 

Low voltage (37.5 V) 

37.500 

37.498 

37.477 

37.521 

Pass 

Low voltage (50 V) 

50.000 

49.999 

49.975 

50.025 

Pass 

Summary of Contents for Keithley 2380-500-30

Page 1: ...Verification Manual 2380 500 30 905 01 Rev A November 2015 P2380 500 30 905 01A 2380 500 30 905 01A www keithley com Models 2380 500 30 Programmable DC Electronic Load Instruments A Greater Measure of...

Page 2: ...photocopy or use of the information herein in whole or in part without the prior written approval of Keithley Instruments is strictly prohibited All Keithley Instruments product names are trademarks...

Page 3: ...be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to...

Page 4: ...y disposed of according to federal state and local laws The WARNING heading in the user documentation explains dangers that might result in personal injury or death Always read the associated informat...

Page 5: ...duction 1 Test Record 2 Performance Verification Procedures 8 Equipment Required 9 Check DC Voltage Setting Accuracy and Voltage Readback Accuracy 11 Check DC Current and DC Current Readback Accuracy...

Page 6: ...also provides a special LED mode to conduct LED electronic load tests by simulating an LED current load Extended warranty Additional years of warranty coverage are available on many products These val...

Page 7: ...ppm rise fall slope Low range 0 0001 0 1 A uS High range 0 001 1 A uS Minimum rising time 10 uS Constant Resistance Operation nominal This value can be set as a constant resistance Resistance Range Lo...

Page 8: ...erform the following steps 1 Remove loose dust on the outside of the electronic load with a lint free cloth and avoid scratching the display 2 Use a soft cloth dampened with water to clean the electro...

Page 9: ...tory Use Electromagnetic Compatibility EMC measurement laboratory and control product family Item 2380 500 30 2380J 500 30 Regional Certifications Classifications and Standards List European Union EC...

Page 10: ...nual before using this product and any associated instrumentation Although some instruments and accessories are used with nonhazardous voltages there are situations where hazardous conditions may occu...

Page 11: ...ne load then any pair of load wires must be capable of safely carrying the full rated short circuit output current of the electronic load Do not loosen any screw on this product There are no user serv...

Page 12: ...00000 0 00149 0 00151 A Pass Low current 0 75 A 0 75000 0 74992 0 74804 0 75179 A Pass Low current 1 5 A 1 50000 1 49994 1 49766 1 50216 A Pass Low current 2 25 A 2 25000 2 24988 2 24723 2 25248 A Pas...

Page 13: ...Low voltage 37 5 V 37 500 37 498 37 466 37 534 V Pass Low voltage 50 V 50 000 49 999 49 963 50 038 V Pass Readback sense accuracy voltage 50 V range Low voltage 0 1 V 0 100 0 099 0 087 0 112 V Pass L...

Page 14: ...0 01 499 76 500 26 V Pass High voltage 500 V 500 00 500 01 499 76 500 26 V Pass Set local accuracy voltage 50 V range Low voltage 0 1 V 0 100 0 100 0 075 0 125 V Pass Low voltage 12 5 V 12 500 12 500...

Page 15: ...e 500 V 500 00 500 01 499 63 500 38 V Pass Readback local accuracy voltage 500 V range High voltage 0 1 V 0 10 0 09 0 03 0 22 V Pass High voltage 125 V 125 00 124 99 124 84 125 15 V Pass High voltage...

Page 16: ...W Pass Power 750 W 750 00 749 57 746 55 752 54 W Pass Set accuracy resistance 10 range Low resistance 0 15 0 150 0 150 0 148 0 152 Pass Low resistance 2 5 2 500 2 500 2 083 3 125 Pass Low resistance...

Page 17: ...pplied in sequence If it is necessary to partially test the DUT applying an individual test a 20 minute warm up period is required for the individual test 3 You must connect the DUT and the test equip...

Page 18: ...than electronic load s rated voltage Keithley 2260B 80 40 3 Digital Multimeter With voltage measuring range from 100 nV to 1 KV and DC resistance measuring range from 100 to 100 M For multi points mea...

Page 19: ...c load RUG Z R001 0 1 TK1 UNR4 3425 OR1BTCRT 7 Current sense resistor cabling Kelvin 4 terminal measuring cables for shunt resistor to voltmeter Guildline 92301 cable set 8 Oscilloscope 20 MHz bandwid...

Page 20: ...thin cable 18 AWG hookup wires 1 Set up the equipment as shown in the following figure Figure 1 Voltage accuracy test setup Ensure the warm up requirement has been met as described in the Performance...

Page 21: ...and 4 then repeat steps from 5 through 8 9 Power off all the equipment after the test 10 Calculate the deviation and determine whether it is within the specification Voltage setting accuracy the measu...

Page 22: ...en the DVM measurement data is stable Record the electronic load s measured current value on VFD displayed value Record the DVM reading value for example the shunt voltage Record the electronic load s...

Page 23: ...y Model 2000 High precision shunt The selected shunt must be able to withstand the maximum current of electronic load Cable for high current thick cable 22 AWG hookup wires Cable for low current thin...

Page 24: ...ment after the test 9 Calculate the deviation and determine whether it is within the specification The measured actual current reading voltage shunt resistance The measured actual power reading voltag...

Page 25: ...Turn on the output for DC source firstly and then turn on the input of electronic load 6 Record the reading values when the DVM measurement data is stable Record the electronic load s measured voltage...

Page 26: ...use of the wire resistance Using remote sensing you can sense the voltage at the electronic load s terminals effectively cancelling the voltage drop caused by the wire resistance When the remote sensi...

Page 27: ...operty of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Cleveland Ohio 44139 44...

Reviews: