Test
description
Expected
value
Measured
value
Low limit
High limit
Unit
Test
result
Readback sense accuracy voltage (500 V range)
High voltage (0.1 V)
0.10
0.09
-0.03
0.22
V
Pass
High voltage (125 V)
125.00
124.99
124.84
125.15
V
Pass
High voltage (250 V)
250.00
249.98
249.80
250.18
V
Pass
High voltage (375 V)
375.00
374.99
374.77
375.20
V
Pass
High voltage (500V)
500.00
500.01
499.76
500.26
V
Pass
High voltage (500 V)
500.00
500.01
499.76
500.26
V
Pass
Set local accuracy voltage (50 V range)
Low voltage (0.1 V)
0.100
0.100
0.075
0.125
V
Pass
Low voltage (12.5 V)
12.500
12.500
12.472
12.528
V
Pass
Low voltage (25 V)
25.000
24.999
24.969
25.031
V
Pass
Low voltage (37.5 V)
37.500
37.499
37.466
37.534
V
Pass
Low voltage (50 V)
50.000
50.000
49.963
50.038
V
Pass
Readback local accuracy voltage (50 V range)
Low voltage (0.1 V)
0.100
0.099
0.087
0.112
V
Pass
Low voltage (12.5 V)
12.500
12.499
12.484
12.515
V
Pass
Low voltage (25 V)
25.000
24.999
24.980
25.018
V
Pass
Low voltage (37.5 V)
37.500
37.498
37.477
37.521
V
Pass
Low voltage (50 V)
50.000
49.999
49.975
50.025
V
Pass