![Tektronix Keithley 2380-500-30 Performance Verification Manual Download Page 15](http://html1.mh-extra.com/html/tektronix/keithley-2380-500-30/keithley-2380-500-30_performance-verification-manual_1077859015.webp)
Model 2380-500-30 Programmable Instruments
Section 3: Performance Verification
2380-500-30-905-01 Rev. A / November 2015
6
Test
description
Expected
value
Measured
value
Low limit
High limit
Unit
Test
result
Set local accuracy voltage (500 V range)
High voltage (0.1 V)
0.10
0.09
-0.15
0.35
V
Pass
High voltage (125 V)
125.00
125.00
124.72
125.28
V
Pass
High voltage (250 V)
250.00
249.99
249.69
250.31
V
Pass
High voltage (375 V)
375.00
374.99
374.66
375.34
V
Pass
High voltage (500 V)
500.00
500.01
499.63
500.38
V
Pass
Readback local accuracy voltage (500 V range)
High voltage (0.1 V)
0.10
0.09
-0.03
0.22
V
Pass
High voltage (125 V)
125.00
124.99
124.84
125.15
V
Pass
High voltage (250 V)
250.00
249.99
249.80
250.18
V
Pass
High voltage (375 V)
375.00
374.99
374.77
375.21
V
Pass
High voltage (500 V)
500.00
500.01
499.76
500.26
V
Pass
High voltage (500 V)
500.00
500.01
499.76
500.26
V
Pass
Set accuracy power (750 W range)
Power (0 W)
0.00
0.00
-1.50
1.50
W
Pass
Power (187.5 W)
187.50
187.42
185.63
189.38
W
Pass
Power (375 W)
375.00
374.77
372.75
377.25
W
Pass
Power (562.5 W)
562.50
562.34
559.88
565.13
W
Pass
Power (750 W)
750.00
749.55
747.00
753.00
W
Pass