Select a measure function
The programmable DC electronic load allows you to perform the
following measure functions.
Measure functions
What the instrument measures
Short function
Simulate a short circuit with an input
Transient mode
Conduct a dynamic response time test for a
power supply
OCP Function
Conduct an automatic test for OCP conditions
OPP Function
Conduct an automatic test for OPP conditions
Battery mode
Conduct a battery capacity test by sinking a
fixed current load
CR-LED mode
Conduct an output current test for an LED
power supply
Measure mode
Conduct a rising/falling time test for a power
supply or a fuse melting time test.
Autotest function Conduct an automatic test with various modes
Front panel key combinations for measurement functions are as
follows:
+
(
Short
)
: Enable/Disable short function.
+
(
Tran
)
: Configure transient parameters
+
(
List
)
: Configure list parameters
+
(
Battery
)
: Configure battery test parameters.
+
(
Program
)
: Configure autotest parameters.
+
(
OCP Test
)
: Configure OCP parameters
+
(
Setup
)
: Configure maximum voltage/current
range for each function
北京海洋兴业科技股份有限公司(证券代码:839145)