Safety
Unpack
Connect
FAQs
Next steps
Introduction
Test
+
(
OPP Test
)
: Configure OPP parameters
Specify a measurement range
You can set value ranges for measurements or allow the instrument
to choose the ranges automatically.
The measurement range determines the input during the
measurement, and also affects the precision of the measurements and
the maximum signal that can be measured.
You can specify the current range and voltage range for the
programmable DC electronic load. The ranges are listed in the next
table.
To set the ranges for CC/CV/CR/CP mode from the
front panel :
1.
Press the
CC/CV/CR/CP
key.
2.
Press
+
.
3.
Set the range of each parameter according to the following table.
4.
Press .
See the "Constant-status operation mode" section of the User
Manual for additional details and the specific parameters
configuration.
Model number
Current measurement
range
Voltage
measurement
range
2380-500-15/
2380J-500-15
3 ~ 15 A
50 ~ 500 V
2380-120-60/
2380J-120-60
6 ~ 60 A
18 ~ 120 V