background image

Safety

Unpack

Connect

FAQs

Next steps

Introduction

Test

8.

Edit the current value for step 2, press

to confirm, the 

following text is displayed.

EDIT  LIST 
Step 002  Level=2A 

9.

Edit the current slope for step 2, press

to confirm, the 

following text is displayed.

EDIT  LIST 
Step 002  Rate=0.1A/uS 

10. Edit the width time for step 2, press

to confirm, the 

following text is displayed.

EDIT  LIST 
Step 002  Width=5S 

11. Edit the list repeating times, press

to confirm, the 

following text is displayed.

EDIT  LIST 
Reapeat Count=3 

12. Save list file and press

to confirm, the following text is 

displayed.

EDIT  LIST 

Save List File=1

1-7

 

13. Press

 , 

select 

On

 and press

to confirm. The 

Trig indicator light turns on. Then press 

to quit the list 

operation when the following text is displayed. 

LIST 
On  Recall  Edit 

14. Press

to turn on the load input. 

15. Press

 

to 

generate a trigger signal. 

16. If you want to quit list operation, press

/

/

/

or any other combination keys. 

 

北京海洋兴业科技股份有限公司(证券代码:839145)

www.hyxyyq.com

Summary of Contents for KEITHLEY 2380-120

Page 1: ...Models 2380 500 and 2380 120 Programmable DC Electronic Load Instruments Quick Start Guide 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 2: ...e personnel are trained to work on live circuits perform safe installations and repair products Only properly trained service personnel may perform installation and service procedures Keithley Instruments products are designed for use with electrical signals that are measurement control and data I O connections with low transient overvoltages and must not be directly connected to mains voltage or ...

Page 3: ...ed in accordance with the operating instructions If the instruments or accessories are used in a manner not specified in the operating instructions the protection provided by the equipment may be impaired Do not exceed the maximum signal levels of the instruments and accessories as defined in the specifications and operating information and as shown on the instrument or test fixture panels or swit...

Page 4: ...clean an instrument remove power from the instrument Use a damp cloth or mild water based cleaner Clean the exterior of the instrument only Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument Products that consist of a circuit board with no case or chassis e g a data acquisition board for installation into a computer should never require cleaning if...

Page 5: ...o 50 KHz The load can be controlled externally from 0 through 10 V using analog interfaces It is equipped with built in ports of RS232 USB and GPIB The load also provides special LED mode to conduct LED power supply test by simulating LED current The models are listed in the table Model number Description 2380 500 15 2380J 500 15 Programmable DC Electronic Load 500V 15A rated for 110 V 230 V nomin...

Page 6: ...on keys 5 Numeric keypad Esc key and combination button 6 Navigation arrow keys Enter Trigger and On Off button 7 Protective cover 8 Input terminals Keypad overview The models 2380 500 15 2380J 500 15 2380 120 60 2380J 120 60 keypad on the front panel is shown in the next figure 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 7: ...ue Sets the load to CR mode and configures the resistance value Sets the load to CP mode and configures the power value Sets the load to trigger mode for list and transient function Turns the instrument on or off Confirms settings Up arrow key Down arrow key Left Right arrow keys Enters numeric values for various parameters Decimal point Cancels the current action and returns to the previous menu ...

Page 8: ...Rear panel description No Description No Description 1 Cooling fan 2 RS232 interface 3 USB interface 4 GPIB interface 5 AC socket fuse inside 6 Line voltage selector 7 I monitor interface 8 Remote sense terminals external trigger and programming input terminals 9 Protective ground terminal 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 9: ...ent immediately You should receive one of the following instruments Model 2380 500 15 2380J 500 15 2380 120 60 2380J 120 60 with the following accessories and documentation Refer to the packing list which comes with the shipment for additional items that are shipped with your instrument No Part Number Description 1 CO XX 161 XXXX XX Power line cord 2 063456100 Programmable DC Electronic Load Produ...

Page 10: ...m is 300 V DC Make sure any test fixture connected to the system protects the operator from contact with hazardous voltages hot surfaces and sharp objects Use shields barriers insulation and safety interlocks to accomplish this Cover the DUT to protect the operator from flying debris in the event of a system or DUT failure Double insulate all electrical connections that an operator can touch Doubl...

Page 11: ...ure proper cooling Do not cover the ventilation holes on the top sides or bottom of the instrument Make sure the instrument is positioned so that it is easy to reach any disconnecting devices such as the power cord and the power switch Power up the instrument The Model 2380 500 15 2380J 500 15 2380 120 60 2380J 120 60 instrument operate at 100 V 115 V 220 V or 230 V with a frequency of 50 Hz or 60...

Page 12: ...ive earth and grounded outlet may result in personal injury or death due to electric shock Do not replace detachable mains supply cords with inadequately rated cords Failure to use properly rated cords may result in personal injury or death due to electric shock To connect line power 1 Turn POWER OFF use front panel power switch 2 Move the line voltage selector on the rear panel to the correct vol...

Page 13: ... table Connection Specifications Front panel input terminals 20 AWG to 12 AWG Sense terminals The wire must be heavy enough not to overheat while carrying the short circuit input current of the unit Make sure to meet the wiring requirements described above 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 14: ...on is not required Keep the wire as short as possible to reduce wire resistance If you want higher measurement precision use four wire remote sense connection To connect the DUT to the load using a two wire connection 1 Unscrew the two nuts and remove the two protective covers as shown in the next figure 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 15: ...2 Connect the cables to the two input terminals as shown in next figure 3 Mount the protective covers to the two input terminals 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 16: ...xt figure 5 Connect the other end of the cables to the DUT The red cable is connected to the positive electrode and the black cable is connected to the negative electrode Failure to install the protective cover may result in personal injury or death due to electric shock 北京海洋兴业科技股份有限公司 证券代码 839145 www hyxyyq com ...

Page 17: ...tive covers to the two input terminals 4 Screw the two input terminals tight 5 Connect the other end of the cables to the DUT The red cable is connected to the positive electrode and the black cable is connected to the negative electrode 6 Connect the DUT to the remote sense terminals on the rear panels as shown in the next wiring diagram Steps1 through 5 above are the same as the steps for two wi...

Page 18: ...creen displays the actual power value and the current voltage power resistance setting values Menu screen overview When you press on the front panel the SYSTEM MENU screen is displayed The following options are displayed on the VFD Initialize Power ON Buzzer Knob Trigger Display Communication and Protocol Select an option by pressing and configure the parameters When you press on the front panel t...

Page 19: ... by sinking a fixed current load CR LED mode Conduct an output current test for an LED power supply Measure mode Conduct a rising falling time test for a power supply or a fuse melting time test Autotest function Conduct an automatic test with various modes Front panel key combinations for measurement functions are as follows Short Enable Disable short function Tran Configure transient parameters ...

Page 20: ...current range and voltage range for the programmable DC electronic load The ranges are listed in the next table To set the ranges for CC CV CR CP mode from the front panel 1 Press the CC CV CR CP key 2 Press 3 Set the range of each parameter according to the following table 4 Press See the Constant status operation mode section of the User Manual for additional details and the specific parameters ...

Page 21: ...Edit Then press to confirm the following text is displayed EDIT LIST High Rate Low Rate 3 Press select High Rate Then press to confirm the following text is displayed EDIT LIST Current Range 3A 4 Edit step count For example press for two steps and then press to confirm the following text is displayed EDIT LIST File Step 2 2 84 5 Edit the current value for step 1 press to confirm the following text...

Page 22: ...ep 002 Width 5S 11 Edit the list repeating times press to confirm the following text is displayed EDIT LIST Reapeat Count 3 12 Save list file and press to confirm the following text is displayed EDIT LIST Save List File 1 1 7 13 Press select On and press to confirm The Trig indicator light turns on Then press to quit the list operation when the following text is displayed LIST On Recall Edit 14 Pr...

Page 23: ...Trig indicator light turns on Press to quit the list operation when the following text is displayed LIST On Recall Edit 4 Press button to turn on the input press to run the list Next steps For more information refer to the Product Information CD ROM which includes the User Manual Quick Start Guide and accessory information These documents provide detailed information about all features of the inst...

Reviews: