Adjustment Procedures
Troubleshooting
Refer to the following table for common issues encountered during the calibration
procedure.
Problem
Things to try
The instrument fails a test
Are the channels hooked up properly, routed to the right channels, and turned on
Is the signal set to the correct waveform, frequency, and termination (if applicable to the test)
Is the output set to
ON
(the red light will be illuminated)
If you make a correction or
fi
nd that everything appears to be correct, click the
Go back a
step
button to try the test again.
I can’t set the amplitude to 0 V, as
speci
fi
ed in the prompt
Ground the signal by pushing the
lower-bezel button.
I can’t
fi
nd the parameters that I
need to set
Is the waveform set correctly
Is the grounding button on? When this is active, you will not be able to make a channel
selection, so in some cases you might need to temporarily deselect the grounding option
while you make a channel selection.
If you can
fi
nd the termination settings, click the
CHANNEL SELECT
lower-bezel button.
I can’t set the voltage to 625 Hz
Set the time marker to 1.6 Ms. This will automatically set the voltage to 625 Hz.
I can’t set the signal to all channels
If you are unable to set the signal to all channels, try setting the signal to Channel 1.
I’m having trouble with tests that
require 6 V or -6 V
Follow the steps for 6 V and -6 V that appear in Step 6 of the DC Voltage test. (See page 3-6,
Ensure that the polarity is correct, by referring to the icon that appears in the upper-left corner
of the calibrator screen.
If the instrument fails a test, go all the way back to the
fi
rst 6 V test (6 V on Channel 1) and
start the sequence over again.
The Sine Signal test failed
If the prompt speci
fi
ed Channel 4, try setting the signal to Channel 5, instead.
3–10
MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual
Summary of Contents for DPO3000 Series
Page 2: ......
Page 14: ...Preface x MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 15: ...Operating Information ...
Page 16: ......
Page 19: ...Theory of Operation ...
Page 20: ......
Page 25: ...Adjustment Procedures ...
Page 26: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 56: ...Maintenance 4 18 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 57: ...Replaceable Parts List ...
Page 58: ......
Page 77: ...Index ...
Page 78: ......