Adjustment Procedures
Time Mark
The Time Mark test is often the second test type to appear in the prompts. When
you receive a time mark prompt, complete the following steps:
1.
Push the
Markers
front-panel button.
2.
Push the
WAVEFORM
lower-bezel button, and then push the
side-bezel button.
NOTE.
If a 625 Hz signal is required, set the time marker to 1.6 ms.
3.
Select the
Amplitude
fi
eld using the front panel scroll buttons, and then
change the voltage by using the large front panel knob on the calibrator.
4.
Ensure that the red light is illuminated to indicate that the output is on. If not,
push the
ON
button.
5.
On the oscilloscope, push the
OK Do Next Step
right-bezel button.
Deskew (MSO Only)
If you are using the MSO3000 Series oscilloscope, and you receive a prompt for a
deskew test, complete the following steps:
1.
Plug the probe, notch down, into the digital channel input, as shown in the
following
fi
gure. Ensure that it snaps into place. Keep the probe plugged in
for the remainder of the calibration procedure.
Figure 3-5: Digital channel input (MSO only)
3–8
MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual
Summary of Contents for DPO3000 Series
Page 2: ......
Page 14: ...Preface x MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 15: ...Operating Information ...
Page 16: ......
Page 19: ...Theory of Operation ...
Page 20: ......
Page 25: ...Adjustment Procedures ...
Page 26: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 56: ...Maintenance 4 18 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 57: ...Replaceable Parts List ...
Page 58: ......
Page 77: ...Index ...
Page 78: ......