Adjustment Procedures
For 0 V
: ground the signal by pushing the grounding
lower-bezel
button. When this is active, you will not be able to make a channel
selection, so in some cases you might need to temporarily deselect the
grounding option while you make a channel selection. If the prompt
requires all channels and the signal is grounded, The calibrator screen
will default to Channel 1.
For any nonzero voltage
: Before you select the voltage, you must
fi
rst
ensure that the calibrator is in direct mode. Push the
lower-bezel
button until 1.0 is selected.
For 6 V or -6 V
:
a.
Push the AUX front panel button.
b.
Push the
side-bezel button for a 6 V signal or push the
side-bezel button for a -6 V signal. The polarity will be indicated only
by the icon that appears in the upper-left corner of the calibrator screen.
c.
Set the amplitude to either 6 V or -6 V.
d.
Set the Pulse/Energy setting to 7.2 Joules (the time should show 10s).
e.
When all of the parameters are set and you ensure that the output is
ON, push the
Trig Pulse
side-bezel and IMMEDIATELY push the
OK Do Next Step
side-bezel button on the oscilloscope.
CAUTION.
Once you push the
Trig Pulse
button on the calibrator,
you have only 9 seconds to push the
OK Do Next Step
button on the
oscilloscope. If the instrument fails one of these tests, go all the way back
to the
fi
rst 6 V test (6 V on channel 1) and start that sequence over again.
7.
Ensure that the red light is illuminated to indicate that the output is on. If not,
push the
ON
button, as shown below:
8.
On the oscilloscope, push the
OK Do Next Step
side-bezel button.
MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual
3–7
Summary of Contents for DPO3000 Series
Page 2: ......
Page 14: ...Preface x MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 15: ...Operating Information ...
Page 16: ......
Page 19: ...Theory of Operation ...
Page 20: ......
Page 25: ...Adjustment Procedures ...
Page 26: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 56: ...Maintenance 4 18 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 57: ...Replaceable Parts List ...
Page 58: ......
Page 77: ...Index ...
Page 78: ......