Adjustment Procedures
Figure 3-4: Factory information dialog
12.
Push the
OK Do Factory Adjustment
side-bezel button to start the
adjustment process. Prompts appear on the oscilloscope screen to indicate the
signal type and the channels to which it should be connected.
NOTE.
The oscilloscope adjusts itself automatically using the calibrator signal as
a reference. You do not need to make any adjustments.
If you make an error, such as connecting the wrong input signal, you can repeat
the last step by pushing the
Return to Previous Screen
side bezel button.
13.
For each step, refer to the next table to identify the signal type, and then
carefully follow the instructions for that signal type on the speci
fi
ed page.
Signal type
Example prompt
Procedure to follow
Parameters to set
DC Voltage
Apply 0 V DC 50
Ω
termination signal to 3.
(See page 3-6,
Waveform type, voltage,
termination, channel
Time Mark
Apply 500 mV Pk-Pk 625 Hz
50
Ω
termination time mark
signal to 3.
(See page 3-8,
Waveform type, voltage,
frequency, termination,
channel
Deskew
Apply 1.0 V Pk-Pk 1 kHz
50
Ω
termination deskew
signal to 3.
(See page 3-8,
Waveform type, voltage,
frequency, termination,
channel
Sine Signal
Apply 2.0 V Pk-Pk 164M
50
Ω
termination sine signal
to 3.
(See page 3-9,
Waveform type, voltage,
frequency, termination,
channel
MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual
3–5
Summary of Contents for DPO3000 Series
Page 2: ......
Page 14: ...Preface x MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 15: ...Operating Information ...
Page 16: ......
Page 19: ...Theory of Operation ...
Page 20: ......
Page 25: ...Adjustment Procedures ...
Page 26: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 56: ...Maintenance 4 18 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 57: ...Replaceable Parts List ...
Page 58: ......
Page 77: ...Index ...
Page 78: ......