Adjustment Procedures
2.
Push the
Aux
front-panel button, and then push the
Zero skew
side-bezel button.
3.
Set the speci
fi
ed voltage, frequency, and termination.
4.
Push the
CHANNEL SELECT
button and ensure that all channels are
selected, including Channel 5.
5.
Ensure that the red light is illuminated to indicate that the output is on. If not,
push the
ON
button.
6.
On the oscilloscope, push the
OK Do Next Step
right-bezel button. In some
cases, this can take several minutes.
CAUTION.
Keep the digital probe plugged in to the oscilloscope. Unplugging the
probe may result in aborting the test.
Sine Signal
When a Sine Signal prompt appears, complete the following steps:
1.
Push the
Sine
front panel button.
2.
To ensure that the voltage is in direct mode, push the
lower-bezel
button and select
1.0
.
3.
Push the
CHANNEL SELECT
lower-bezel button, and then ensure that
Channel 5 is selected.
4.
Ensure that the red light is illuminated to indicate that the output is on. If not,
push the
ON
button.
5.
On the oscilloscope, push the
OK Do Next Step
right-bezel button.
MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual
3–9
Summary of Contents for DPO3000 Series
Page 2: ......
Page 14: ...Preface x MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 15: ...Operating Information ...
Page 16: ......
Page 19: ...Theory of Operation ...
Page 20: ......
Page 25: ...Adjustment Procedures ...
Page 26: ......
Page 37: ...Maintenance ...
Page 38: ......
Page 56: ...Maintenance 4 18 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Page 57: ...Replaceable Parts List ...
Page 58: ......
Page 77: ...Index ...
Page 78: ......