PS2520 Series Performance Verification
46
Handheld and Benchtop Instruments Basic Service
Complete the following procedures to verify constant voltage source accuracy.
Outputs 1 and 2.
Use the following steps to check the OUTPUT 1 and
OUTPUT 2 accuracy.
1. Set up the Variac. Connect the power supply AC input to the Variac variable
output. Allow 20 minutes for the power supply to warm up and stabilize.
2. Ensure that the power supply output is disabled. Connect the DMM and
electronic load to the front panel OUTPUT 1 terminals. See Figure 3 for
details.
+
–
Power supply
Electronic load
–
+
Digital multimeter
+
–
10 A
COM
Figure 8: Constant Voltage Source Test Setup
3. Set the DMM to measure 5 amperes DC.
4. Press SHIFT
→
OUT 1; verify that the “1” indicator lights up on the display.
5. Set up the power supply as follows:
VOLTS SET
20 V
CURRENT SET
2.55 A
OVP SET
22 V
6. Press OUTPUT ON/OFF. Verify that the “OUT” indicator lights up on the
display.
7. Enable the electronic load. Adjust the load until the DMM current reads
2.500 A.
8. Adjust the output of the Variac from 108 to 132 VAC (120 V range) or 198
to 242 VAC (220 V range). Verify that the power supply VOLTS readout
changes
≤
0.003 volts over the adjustment range.
9. Press OUTPUT ON/OFF and disable the electronic load.
10. Press SHIFT
→
OUT 2; verify that the “2” indicator lights up on the display.
Constant Voltage
Source Accuracy
Summary of Contents for DCM300
Page 10: ...Preventing Electrostatic Discharge viii Handheld and Benchtop Instruments Basic Service ...
Page 13: ...Instructions Manual DCM300 and DCM320 Digital Clamp Multimeters 070 9847 00 ...
Page 15: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 30: ...Instructions Manual DCM330 Digital Clamp Meter 070 9848 00 ...
Page 32: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 49: ...Instructions Manual DCM910 Digital Clamp Meter 070 9849 00 ...
Page 51: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 68: ...Instructions Manual DMM800 Series Digital Multimeters 070 9850 00 ...
Page 70: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 110: ...DMM800 Series Adjustment Procedures 40 Handheld and Benchtop Instruments Basic Service ...
Page 111: ...Instructions Manual DMM912 914 and 916 Digital Multimeters 070 9851 00 ...
Page 113: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 153: ...DMM91X Series Adjustment Procedures 40 Handheld and Benchtop Instruments Basic Service ...
Page 154: ...Instructions Manual DTM500 Series Digital Thermometers 070 9852 00 ...
Page 156: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 173: ...Instructions Manual DTM900 and DTM920 Digital Thermometers 070 9853 00 ...
Page 175: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 192: ...Instructions Manual PS280 and PS283 Power Supplies 070 9894 00 ...
Page 194: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 217: ...Instructions Manual PS2520 Series Power Supplies 070 9854 00 ...
Page 219: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 286: ...Instructions Manual TAS 200 Series Oscilloscopes 070 9855 00 ...
Page 288: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 332: ...TAS 200 Series Adjustment Procedures 44 Handheld and Benchtop Instruments Basic Service ...
Page 333: ...Instructions Manual THM420 Digital Multimeter 070 9856 00 ...
Page 335: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 353: ...THM420 Adjustment Procedures 18 Handheld and Benchtop Instruments Basic Service ...
Page 354: ...Instructions Manual THM500 Series Instruments 070 9857 00 ...
Page 356: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 380: ...THM500 Series Performance Verification 24 Handheld and Benchtop Instruments Basic Service ...
Page 381: ......
Page 382: ......