DCM330 Adjustment Procedures
12
Handheld and Benchtop Instruments Basic Service
Preparation for Adjustment
The following guidelines apply to all DCM330 adjustments.
H
Perform all adjustments in a 21
_
to 25
_
C ambient environment with a
relative humidity of 75% or less.
H
Before making any adjustment, warm up the current meter for at least
30 minutes.
H
Do not alter any setting without reading the entire adjustment procedure first.
H
Do not alter a setting unless a performance characteristic cannot be met at
the current setting.
H
Read the Safety Summary at the beginning of this manual.
You must open the meter case to gain access to the internal adjustments.
1. Lay the meter face down on a flat work surface.
2. Remove the two screws from the case bottom with a Phillips-head
screwdriver.
3. Gently lift the end of the case bottom until it unsnaps from the case top.
4. Remove the three screws that secure the circuit board assembly to the case
top. Do not remove the screws that secure the circuit boards to each other.
5. To access the adjustments, lift the circuit board assembly far enough out of
the top case to expose the adjustments. See Figure 2 and the procedure that
follows.
To reassemble the meter following the adjustments, perform steps 2 through 4
above in reverse order.
Adjustment Procedure
To return your instrument to factory calibration, implement the following
procedures.
Use a small flat-tipped screwdriver to make the adjustments. Refer to Figure 2
for adjustment locations.
WARNING.
Magnetic fields are produced that may cause a malfunction in heart
pacemakers, or damage to sensitive equipment.
Open the Meter Case
Summary of Contents for DCM300
Page 10: ...Preventing Electrostatic Discharge viii Handheld and Benchtop Instruments Basic Service ...
Page 13: ...Instructions Manual DCM300 and DCM320 Digital Clamp Multimeters 070 9847 00 ...
Page 15: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 30: ...Instructions Manual DCM330 Digital Clamp Meter 070 9848 00 ...
Page 32: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 49: ...Instructions Manual DCM910 Digital Clamp Meter 070 9849 00 ...
Page 51: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 68: ...Instructions Manual DMM800 Series Digital Multimeters 070 9850 00 ...
Page 70: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 110: ...DMM800 Series Adjustment Procedures 40 Handheld and Benchtop Instruments Basic Service ...
Page 111: ...Instructions Manual DMM912 914 and 916 Digital Multimeters 070 9851 00 ...
Page 113: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 153: ...DMM91X Series Adjustment Procedures 40 Handheld and Benchtop Instruments Basic Service ...
Page 154: ...Instructions Manual DTM500 Series Digital Thermometers 070 9852 00 ...
Page 156: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 173: ...Instructions Manual DTM900 and DTM920 Digital Thermometers 070 9853 00 ...
Page 175: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 192: ...Instructions Manual PS280 and PS283 Power Supplies 070 9894 00 ...
Page 194: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 217: ...Instructions Manual PS2520 Series Power Supplies 070 9854 00 ...
Page 219: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 286: ...Instructions Manual TAS 200 Series Oscilloscopes 070 9855 00 ...
Page 288: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 332: ...TAS 200 Series Adjustment Procedures 44 Handheld and Benchtop Instruments Basic Service ...
Page 333: ...Instructions Manual THM420 Digital Multimeter 070 9856 00 ...
Page 335: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 353: ...THM420 Adjustment Procedures 18 Handheld and Benchtop Instruments Basic Service ...
Page 354: ...Instructions Manual THM500 Series Instruments 070 9857 00 ...
Page 356: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 380: ...THM500 Series Performance Verification 24 Handheld and Benchtop Instruments Basic Service ...
Page 381: ......
Page 382: ......