TAS 200 Series Performance Verification
14
Handheld and Benchtop Instruments Basic Service
Horizontal Check
The following checks verify the horizontal accuracy of your oscilloscope.
To check time base accuracy, perform the following steps.
1. Use the 50
W
precision coaxial cable to connect the output of the time mark
generator to the 50
W
termination; then connect the 50
W
termination to the
TAS 200 series instrument CH 1 input. See Figure 4 below.
Time Mark
Generator
50
W
Termination
Precision cable
Figure 4: Timing Check Setup
2. Set up the oscilloscope as follows:
VERTICAL MODE
CH1
CH 1 AC-DC
DC
CH 1 VOLTS/DIV
0.5 V
CH 1 GND
Out (release)
HORIZONTAL SEC/DIV
.1
m
s
TRIGGER MODE
AUTO
TRIGGER COUPLING
DC
3. Set up the time mark generator to produce 0.1
m
s markers.
4. Center the time marks vertically on the oscilloscope display.
5. Position the rising edge of the second time mark to the second vertical
graticule line of the oscilloscope display.
6. Verify that the time mark to graticule accuracy over the center eight divisions
is within the limits shown for each HORIZONTAL SEC/DIV setting listed
in Table 15.
Time Base Accuracy
Summary of Contents for DCM300
Page 10: ...Preventing Electrostatic Discharge viii Handheld and Benchtop Instruments Basic Service ...
Page 13: ...Instructions Manual DCM300 and DCM320 Digital Clamp Multimeters 070 9847 00 ...
Page 15: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 30: ...Instructions Manual DCM330 Digital Clamp Meter 070 9848 00 ...
Page 32: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 49: ...Instructions Manual DCM910 Digital Clamp Meter 070 9849 00 ...
Page 51: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 68: ...Instructions Manual DMM800 Series Digital Multimeters 070 9850 00 ...
Page 70: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 110: ...DMM800 Series Adjustment Procedures 40 Handheld and Benchtop Instruments Basic Service ...
Page 111: ...Instructions Manual DMM912 914 and 916 Digital Multimeters 070 9851 00 ...
Page 113: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 153: ...DMM91X Series Adjustment Procedures 40 Handheld and Benchtop Instruments Basic Service ...
Page 154: ...Instructions Manual DTM500 Series Digital Thermometers 070 9852 00 ...
Page 156: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 173: ...Instructions Manual DTM900 and DTM920 Digital Thermometers 070 9853 00 ...
Page 175: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 192: ...Instructions Manual PS280 and PS283 Power Supplies 070 9894 00 ...
Page 194: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 217: ...Instructions Manual PS2520 Series Power Supplies 070 9854 00 ...
Page 219: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 286: ...Instructions Manual TAS 200 Series Oscilloscopes 070 9855 00 ...
Page 288: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 332: ...TAS 200 Series Adjustment Procedures 44 Handheld and Benchtop Instruments Basic Service ...
Page 333: ...Instructions Manual THM420 Digital Multimeter 070 9856 00 ...
Page 335: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 353: ...THM420 Adjustment Procedures 18 Handheld and Benchtop Instruments Basic Service ...
Page 354: ...Instructions Manual THM500 Series Instruments 070 9857 00 ...
Page 356: ...Table of Contents ii Handheld and Benchtop Instruments Basic Service ...
Page 380: ...THM500 Series Performance Verification 24 Handheld and Benchtop Instruments Basic Service ...
Page 381: ......
Page 382: ......