In the dc mode, when the triggering signal is
obtained from either the lower or the upper
beam amplifiers, varying the respective vertical
position controls will change the triggering point.
For this reason, you may find it necessary to re
adjust the TRIGGERING LEVEL control when you
change the vertical position of the trace. If you
desire to eliminate this effect, you can use the
ac mode providing the triggering signal is other
wise suitable for this mode of operation. In the
dc mode, the dc level of external triggering
signals will also effect the triggering point. Gen
erally, when the triggering signal is small com
pared to its dc level, the ac mode should be
used.
AC Mode
The ac mode is selected in the LINE, LOWER
AC, UPPER AC, and EXT AC positions of the
TRIGGER SELECTOR switch. This mode provides
stable triggering on virtually all types of wave
forms. As a general rule, however, the ac mode
is unsatisfactory for triggering with low ampli
tude waveforms at frequencies below approxi
mately 15 cycles. This figure will vary depend
ing upon the amplitude and shape of the trig
gering waveform and should not therefore be
set as an absolute standard. Triggering at fre
quencies below 15 cycles can be accomplished
when higher amplitude triggering signals are
used.
In the ac mode, the triggering point depends
on the average voltage level of the triggering
signals. If the triggering signals occur at ran
dom, the average voltage level will vary causing
the triggering point to also vary. This shift of
the triggering point may be enough so that it is
impossible to maintain a stable display. In such
cases you should use the dc mode.
How to select triggering slope.
If you wish to trigger the sweep on the rising
(positive slope) portion of the triggering wave
form, place the TRIGGER SELECTOR switch at
+ . If you wish to trigger on the falling (nega
tive slope) portion of the triggering waveform,
place the TRIGGERING SELECTOR switch at —.
How to set the triggering level.
In the ac and dc triggering modes, the TRIG
GERING LEVEL control determines at which point
on the triggering signal that the sweep is trig
gered. Using this control, the sweep can be con
tinuously triggered at any point on the waveform
so long as the slope of the waveform is great
enough to provide stable triggering. In the DC
mode, the sweep cannot be triggered with any
degree of stability at the top of a square wave,
for example, because the time that the voltage
remains constant is comparatively long. As a
result, the sweep triggers at random points along
the top of the square wave producing consider
able trace jitter.
You can use the same method to set the TRIG
GERING LEVEL control for either the ac or dc
mode. After selecting the triggering slope, ro
tate the TRIGGERING LEVEL control fully count
erclockwise to the AUTOMATIC position. Then
rotate the TRIGGERING LEVEL control clockwise
until the sweep no longer triggers. Continue to
rotate the control in the clockwise direction until
the sweep again triggers and a stable display is
obtained. Further rotation of the control in the
clockwise direction causes the sweep to trigger
at more positive points on the triggering wave
form. See Figure 3-4.
FREE-RUNNING OPERATION
With the Type 502, you can get a periodic,
free-running sweep, independent of any external
triggering or synchronizing signal by rotating the
TRIGGERING LEVEL control fully clockwise to the
RECURRENT position. This permits you to ob
serve the upper- and lower-beam traces without
an input signal.
OPERATING INSTRUCTIONS — TYPE 502
3-5
Summary of Contents for 502 series
Page 2: ......
Page 6: ......
Page 7: ...Fig 1 1 Type 502 Oscilloscope with accessories ...
Page 8: ......
Page 36: ......
Page 50: ......
Page 77: ...0 0 TYPE 502 OSCILLOSCOPE A BLOCK DIAGRAM ...
Page 79: ...INPUT A M PLIFIE R SECOND D R IV E R O UTPUT A M PLIFIE R C F S AM PLIFIE R X 10 X 100 ...
Page 82: ...V722 type 502 o s c illo s c o p e HEATER WIRING DIAGRAM A POWER SUPPLY ...
Page 90: ...M ILLE RRUNUP CIRCU IT 350 V 2 3 M ILUSEC TIME BASE GENERATOR 12 8 58 ...
Page 94: ...CALIBRATOR MULTIVIBRATOR TYPE 502 OSCILLOSCOPE A ...