Performance Check/Calibration—Type 310A
between the second and tenth vertical lines of the graticule.
The marker at the tenth vertical line must be within 0.4
division (within 5%) of the line when the marker at the
second vertical line is positioned exactly.
f.
ADJUST—C250 (see Fig. 5-18) for one marker each
division (use low-capacitance screwdriver). The markers at
the second and tenth vertical lines must coincide exactly
with their respective graticule lines (reposition the display
with the H O R IZO N TAL POSITION control if necessary).
36. Check Normal Sweep Timing Accuracy
CAUTION
To prevent burning of the CRT phosphor at slow
sweep rates, position the baseline of the marker dis
play below the viewing area.
a. Change the following control settings
TRIGGER
+ INT
Horizontal Display
X I MAG
b. CHECK-Using the T IM E /D IV switch and time-mark
generator settings given in Table 5-4, check normal sweep
timing within 0.24 division (within 3%) over the center
eight divisions of the display. Set the TRIG LEVEL control
as necessary for a stable display.
TABLE 5-4
Normal Sweep Timing Accuracy
T IM E /D IV switch
Time mark
CRT display
setting
generator output
(markers/division)
11 M' T 1
0.5 microsecond
1
1
C
1 microsecond
1
2 pSEC
1 microsecond
2
5 pSEC
5 microsecond
1
10/
j
SEC
10 microsecond
1
20 pSEC
10 microsecond
2
bOpSEC
50 microsecond
1
.1 MSEC
0.1 millisecond
1
.2 MSEC
0.1 millisecond
2
.5 MSEC
0.5 millisecond
1
1 MSEC
1 millisecond
1
2 MSEC
1 millisecond
2
5 MSEC
5 millisecond
1
10 MSEC
10 millisecond
1
20 MSEC
10 millisecond
2
50 MSEC
50 millisecond
1
.1 SEC
0.1 second
1
.2 SEC
0.1 second
2
37. Check Magnified Sweep Timing Accuracy
a. Change the following control settings:
TRIGGER
+ EXT
Horizontal Display
X5 MAG
b. Set the time-mark generator for one-microsecond
trigger output.
®
Fig. 5 -1 8 . Lo cation o f high-fraquancy adjustm ents (la ft sida).
c. Connect the time-mark generator trigger output con
nector to the EXT TRIG INPUT binding post with the
18-inch BNC-to banana plug patch cord.
d. CHECK-Using the T IM E /D IV switch and time-mark
generator settings given in Table 5-5, check magnified
sweep timing within the given tolerances over the center
eight divisions of the total magnified sweep length. Set the
TRIG LEVEL control as nocessary for a stable display and
set the TRIGGER switch for + INT source as given in Table
5-5.
e. Disconnect all test equipment.
TABLE 5-5
Magnified Sweep Timing Accuracy
T IM E /D IV
switch
setting
Time-mark
generator
output
CRT display
(markers/
division)
Allowable
error for
given
accuracy
,5pSEC
.1 microsecond
1
0.4/livision
(within 5%)
IpSEC
.1 microsecond
2
0.32 division
(within 4%)
0
2pSEC
.1 microsecond
4
Set TRIGG
R switch to + 1NT
bpSEC
1 microsecond
1
lOjtSEC
1 microsecond
~ T ~
20 pSEd
1 microsecond
4
50/uSEC
10 microsecond
1
1
MSEC
10 microsecond
2
.2 MSEC
10 microsecond
4
.5 MSEC
.1 millisecond
1
1 MSEC
.1 millisecond
2
2 MSEC
.1 millisecond
4
5 MSEC
1 millisecond
1
10 MSEC
1 millisecond
2
20 MSEC
1 millisecond
4
50 MSEC
10 millisecond
1
.1 SEC
10 millisecond
2
.2 SEC
10 millisecond
4
5-23
Summary of Contents for 310A
Page 3: ...Type 310A ...
Page 8: ...NOTES ...
Page 10: ...Operating Instructions Type 310A 2 2 Fifl 2 2 Front and rear panel controls and connectors ...
Page 16: ...rsj 60 T Y P E 3 1 0 A T E S T S E T U P C H A R T Fig 2 3 Operating Instructions Type 310A ...
Page 28: ...NOTES ...
Page 44: ...NOTES ...
Page 88: ...SECTION 7 M ECHANICAL PARTS LIST Type 310A ...
Page 98: ...Mechanical Parts List Type 310A POWER CHASSIS 7 1 0 ...
Page 102: ...Mechanical Parts List Type 310A VERTICAL AMPLIFIER CHASSIS 7 1 4 ...
Page 106: ...Mechanical Parts List Type 310A REAR 7 1 8 ...
Page 113: ...V 8 59 B L O C K D I A G R A M ...
Page 116: ...0 ...
Page 119: ... 1 TIM E DIV 1 SW 75C SWI75D I t ...
Page 129: ... CRT CIRCUIT C IR C U IT NUMBfcRS 7 0 0 THRU 7 6 0 ...
Page 131: ...CALIB R A TO R CATHODE FO LLO W E R IOOV CALIBRATOR c C IR C U IT NUMBfep S SOO THRU 5 4 9 ...