Performance Check/Calibration-Type 310A
25. Check Low-Frequency Triggering Operation
a. Change the following control settings:
VO LTS/D IV
0.2
TRIGGER
+ IN T
Trigger Coupling-Mode
AUTO
T IM E /D IV
1 MSEC
b. Connect the low-frequency constant-amplitude sine-
wave generator to the vertical INPUT connector through
the BNC-to-dual binding post adapter, 42-inch 50-ohm BNC
cable, 50-ohm termination and the BNCT connector. Con
nect the output of the BNC T connector to the EXT TRIG
INPUT binding post with the 18-inch BNC-to-banana plug
patch cord.
c. Set the low-frequency generator for a 0.25-division
display at one kilohertz.
d. CHECK—Stable CRT display can be obtained with
the Trigger Coupling-Mode switch set to AUTO, AC and DC
(TRIG LEVEL control may be adjusted as necessary to
obtain a stable display).
e. Set the TRIGGER switch to + EXT.
f. Set the low-frequency generator for a one-division dis
play (0.2 volt) at one kilohertz.
g. CHECK—Stable CRT display can be obtained with
the Trigger Coupling-Mode switch set to AUTO, AC and DC
(TRIG LEVEL control may be adjusted as necessary to
obtain a stable display).
h. Set the T IM E /D IV switch to 10 MSEC.
i. Set the low-frequency constant-amplitude generator
for an output frequency of 60 hertz (output amplitude
remains at 0.2 volt).
j. Set the Trigger Coupling-Mode switch to AUTO.
k. CHECK—Stable CRT display is presented.
l
. Set the low-frequency constant-amplitude generator
for an output frequency of 30 hertz (output amplitude
remains at 0.2 volt).
m. CHECK-Stable CRT display can be obtained with
the Trigger Coupling-Mode switch set to AC and DC (TRIG
LEVEL control may be adjusted as necessary to obtain a
stable display).
n. Set the TRIGGER switch to + INT.
o. Set the low-frequency generator for a 0.25-division
display at 30 hertz.
p. CHECK—Stable CRT display can be obtained with
the Trigger Coupling-Mode switch set to AC and DC (TRIG
LEVEL control may be adjusted as necessary to obtain a
stable display).
q. Set the Trigger Coupling-Mode switch to AUTO.
r. Set the low-frequency generator for a 0.25-division
display at 60 hertz.
s. CHECK—Stable CRT display is presented.
26. Check Triggering Slope Operation
a. Set the T IM E /D IV switch to 200 ^SEC.
b. Set the low-frequency generator fo r a four-division
display at one kilohertz.
c. CHECK—CRT display starts on the positive slope of
the waveform.
d. Set the TRIGGER switch to — INT.
e. C H EC K-C R T display starts on the negative slope of
the waveform.
f. Set the TRIGGER switch to + EXT.
g. C H EC K-C R T display starts on the positive slope o f
the waveform.
h. Set the TRIGGER switch to — EXT.
i. C H EC K-C R T display starts on the negative slope o f
the waveform.
j. Disconnect all test equipment.
27. Check Line Triggering Operation
a. Connect the 10X probe to the vertical INPUT con
nector.
b. Change the following control settings:
VO LTS/DIV
10
TRIGGER
+ LINE
T IM E /D IV
2 MSEC
c. Connect the probe tip to the same line-voltage source
which is connected to this instrument.
d. CHECK-Stable CRT display triggered on the positive
slope.
e. Set the TRIGGER switch to — LINE.
f. CHECK-Stable CRT display triggered on the negative
slope.
g. Disconnect all test equipment.
28. Check/Adjust Normal Timing
a. Change the follow ing control settings:
5-20
Summary of Contents for 310A
Page 3: ...Type 310A ...
Page 8: ...NOTES ...
Page 10: ...Operating Instructions Type 310A 2 2 Fifl 2 2 Front and rear panel controls and connectors ...
Page 16: ...rsj 60 T Y P E 3 1 0 A T E S T S E T U P C H A R T Fig 2 3 Operating Instructions Type 310A ...
Page 28: ...NOTES ...
Page 44: ...NOTES ...
Page 88: ...SECTION 7 M ECHANICAL PARTS LIST Type 310A ...
Page 98: ...Mechanical Parts List Type 310A POWER CHASSIS 7 1 0 ...
Page 102: ...Mechanical Parts List Type 310A VERTICAL AMPLIFIER CHASSIS 7 1 4 ...
Page 106: ...Mechanical Parts List Type 310A REAR 7 1 8 ...
Page 113: ...V 8 59 B L O C K D I A G R A M ...
Page 116: ...0 ...
Page 119: ... 1 TIM E DIV 1 SW 75C SWI75D I t ...
Page 129: ... CRT CIRCUIT C IR C U IT NUMBfcRS 7 0 0 THRU 7 6 0 ...
Page 131: ...CALIB R A TO R CATHODE FO LLO W E R IOOV CALIBRATOR c C IR C U IT NUMBfep S SOO THRU 5 4 9 ...