Applications
More accurate phase measurements can be made by using the
horizontal magnifier function to increase the sweep speed with
out changing the SEC/DIV variable control setting. To do this,
set the HORIZONTAL MAG to X I 0.
With the sweep speed increased 10 times (MAG set to X I 0), the
magnified horizontal graticule calibration will be 4 .5 ° per division
(45°/division divided by 10). Figure 3 -9 shows the same signals
illustrated in Figure 3 -8 , but horizontally magnified by a factor of
10
.
EXAMPLE. In Figure 3 -9 , the 10X magnified display results in a
horizontal difference of six divisions between the two signals.
Substituting the given values into the phase difference formula:
Phase Difference = 6 div x 4 .5 °/d iv = 27°.
CHANNEL 1 (REFERENCE)
CHANNEL 2 (COMPARISON)
MEASURE
TIME FROM
A T O B
6 7 1 7 -
12
Figure 3 -9 . High-resolution phase difference.
3-16
2205 Operator's
Summary of Contents for 2205
Page 1: ......
Page 14: ...SECTION 1 PREPARATION FOR USE 2205 Operator s ...
Page 15: ...1 ...
Page 21: ......
Page 22: ...SECTION 2 OPERATION 2 2205 Operator s ...
Page 23: ... vV V vi_rf M r ...
Page 24: ...NOTE Numbers on the instrument are keyed to Table 2 1 ...
Page 39: ...SECTION 3 APPLICATIONS 2205 Operator s ...
Page 40: ......
Page 60: ...F M ...
Page 61: ...SECTION 4 CHECKS AND ADJUSTMENTS 2205 Operator s ...
Page 62: ...B 3 ...
Page 66: ......
Page 67: ...SECTION 5 PERFORMANCE CHARACTERISTICS 2205 Operator s ...
Page 68: ...5 ...
Page 77: ...SECTION 6 OPTIONS AND ACCESSORIES ...
Page 78: ...6 ...
Page 86: ......
Page 88: ......