Applications
Substituting given values in the formula:
Rise Time
5 div x 1 us/div
1
5. p.s
Tim e Difference Between Pulses on T im e-R elated Signals
The calibrated sweep speed and dual-trace features of the 2205
allow measurement of the time difference between two separate
events. To m easure.tim e difference, use the following proce
dure:
1. Set the TRIGGER SOURCE to C H I.
2. Set both input coupling controls to the same position.
3. Using probes or cables with equal time delays, apply a
known reference signal to the CHI input connector and ap
ply the comparison signal to the CH2 input.
4. Set both VOLTS/DIV controls for a display of about five divi
sions.
5. Set VERTICAL MODE to BOTH; then select either ALT or
CHOP, depending on the frequency of the input signals.
6. If the two signals are opposite in polarity, set the middle
VERTICAL MODE selector to CH2 INVERT to invert the Chan
nel 2 display.
7. Adjust the TRIGGER LEVEL control for a stable display.
8. Set the SEC/DIV control to a sweep speed that provides
three or more divisions of horizontal separation between
measurement points on the two displays. Center each of
the displays vertically (see Figure 3 -7 ).
3-12
2205 Operator's
Summary of Contents for 2205
Page 1: ......
Page 14: ...SECTION 1 PREPARATION FOR USE 2205 Operator s ...
Page 15: ...1 ...
Page 21: ......
Page 22: ...SECTION 2 OPERATION 2 2205 Operator s ...
Page 23: ... vV V vi_rf M r ...
Page 24: ...NOTE Numbers on the instrument are keyed to Table 2 1 ...
Page 39: ...SECTION 3 APPLICATIONS 2205 Operator s ...
Page 40: ......
Page 60: ...F M ...
Page 61: ...SECTION 4 CHECKS AND ADJUSTMENTS 2205 Operator s ...
Page 62: ...B 3 ...
Page 66: ......
Page 67: ...SECTION 5 PERFORMANCE CHARACTERISTICS 2205 Operator s ...
Page 68: ...5 ...
Page 77: ...SECTION 6 OPTIONS AND ACCESSORIES ...
Page 78: ...6 ...
Page 86: ......
Page 88: ......