Applications
CHANNEL 1 (REFERENCE)
CHANNEL 2
6 7 1 7 - 1 0
Figure 3 -7 . Time difference between pulses on
tim e-related signals.
9. Determine the horizontal difference between the two signal
measurement points and calculate the time difference using
the following formula:
horizontal
SEC/DIV
difference
x
control
Time
_
(divisions)
setting
Difference
------------ ------------- ;-------------
magnification factor
EXAMPLE. In Figure 3 -7 , the. SEC/DIV control is set to 50
jj
.
s
per division and HORIZONTAL MAG is set to X I 0. The horizontal
difference between waveform measurement points is 4 .5 divi
sions.
Time
Difference
50 us/div x 4 .5 div
10
22.5 jis
2205 Operator’s