Applications
6. Adjust the TRIGGER LEVEL control to obtain a stable display.
7. Set the SEC/DIV control to a position that displays at least
one cycle of the signal.
8. Measure the divisions of vertical deflection between the
ground reference line and the point on the waveform at
which the level is to be determined (see Figure 3 -2 ).
l e v e l
1(
A
\ '
0
.
/
ZERO-VOLT
REFERENCE
i
0
/
LEVEL
1
B
6717-05
Figure 3 -2 .
Instantaneous voltage measurement.
9. Calculate the instantaneous voltage, using the following for-
mula:
Instan
vertical
polarity
VOLTS/DIV
taneous
=
deflection
x
(+ or - )
x
setting
Voltage
(divisions)
(10X Probe)
* lf the signal adapter is being used for the measurement,
use the IX VOLTS/DIV setting.
3-4