Calibration— Type 1A1
15 -p l Inpul
Copocitanc*
N orm o liitr
Typ« 106
Squar»-Wove Generator
10 X Allqnwol ©r
50 ohm
In-lino
Termination
J
OR-lo-BNC
Adapter
S-ni
50-ohm
Cable
Rig. 7-24. Tail equipment letup tor flop 21.
Control Settings
Type 547 Oscilloscope
Horizontal Display
B
Sweep Magnifier
X Off
Single Sweep Switch
Normal
Triggering Level
Near 0 and pushed in
Triggering Source
Norm Inf
Triggering Coupling
AC
Triggering Slope
+
Triggering Mode
Auto Stability
Time/Cm
.5 msec
Horizontal Position
Trace positioned to stc
near left graticule lii
Amplitude Calibrator
Off
Type 1A1
MODE
CH 1
POSITION
(both channels)
Near midrange
PULL FOR INVERT
(both channels)
Normal or in
VARIABLE VOLTS/CM
(both channels)
CALIB
VOLTS/CM
(both channels)
.05
INPUT SELECTOR
(both channels)
DC
Type 106 Square-Wave Generator
Repetition Rote Range
Multiplier
Symmetry
Amplitude
Hi Amplitude Fast Rise
switch
Fast Rise controls
Power
1 kHz
2.5
Midrange
Fully CCW
Hi Amplitude
Not applicable
On
21. Adjust Input Capacitance and
O
Attenuator Compensation (Both Channels)
a. Test equipment setup, with connections made at com
pletion of step 21c, is shown in Fig. 7-24.
b. Set the Type 106 Square-Wave Generator control to
the positions given in the list that precedes step 21.
c. Apply the 2.5 kHz signal from the Type 106 high am
plitude output connector through a 10X attenuator, 5-ns
50-ohm cable, 50-ohm in-line termination and a 15-pF input
capacitance normalizer to the Channel 1 INPUT connector
on the Type 1A1 (see Fig. 7-24).
7-22
© I