Calibration— Type 1A1
e. Using the Channel 2 POSITION control, position the
trace to coincide with graticule center. Leave the control
at this setting until completing the next step.
f. INTERACTION— Interacts with step 4.
4. Adjust Channel 2 .05 VOLTS DC BAL
O
(R248)
a. Equipment setup is shown in Fig. 7-2.
b. Set the Channel 2 VOLTS/CM switch to .05.
c. CHECK— That the trace coincides with graticule cen
ter.
d. ADJUST— Channel 2 .05 VOLTS DC BAL control (see
Fig. 7-6) to vertically position the trace to coincide with
graticule center.
e. INTERACTION— Interacts with step 3.
7-8