Calibration— Type 1A1
\
4? Inch, SO ohm
Cable
Fig. 7 -3 3 , Toct equipment itlu p for ito p 30.
Control Settings
Type 547 Oscilloscope
Horizontal Display
Swoop Magnifier
Single Swoop Swilch
Triggering Lovel
Triggering Source
Triggering Coupling
Triggering Slope
Triggering Mode
Time/Cm
Variable (Time/Cm)
Horizontal Position
Amplitude Calibrator
Type
MODE
POSITION
(both channels)
PULL FOR INVERT
(both channels)
VARIABLE VOLTS/CM
(both channels)
VOLTS/CM Chan
nel 1)
B
X I Off
Normol
Near 0 and pushed in
Norm lot
AC
+
Auto Stability
.5 msec
Calibrated
Trace positioned to
start at left graticule
line
Off
1A1
CH 1
Midrange
Normal or in
CALIB
.005
VOLTS/CM (Chan
nel 21
.005
INPUT SELECTOR
(both channels)
DC
Standard Amplitude Calibrator
Amplitude
Mode
Mixed
X 100 Amplifcr
Power
20 mVolts
Square
Up
Not Applicable
On
20. Check VOLTS/CM Attenuation Ratios
(Both Channels)
a. Test equipment setup, with connections made at com
pletion of step 20b, is shown in Fig. 7-23.
b. Connect a 42-inch 50-ohm cable from Channel 1 IN
PUT connector to the Standard Amplitude Calibrator out
put connector.
c. Check that the Standard Amplitude Calibrator is set
for an output of 20 mV.
d. Adjust the oscilloscope Triggering Level control to
obtain o stable display.
e. Using the Channel 1 POSITION control, align the dis
play with the graticule lines so the amplitude con be meas
ured easily.
7-20