Appendix D: Performance Verification
D–4
THS710A, THS720A, THS730A & THS720P User Manual
Performance Verification Procedures
Before beginning these procedures, two conditions must first be met:
H
The TekScope instrument must have been operating continuously
for ten minutes in an 18
°
C to 28
°
C ambient environment with a
relative humidity of less than 60%.
H
You must perform the Compensate Signal Path operation
described on page 2–11. If the operating temperature changes by
more than 5
°
C, you must perform the Compensate Signal Path
operation again.
The time required to complete the entire procedure is approximately
one hour.
WARNING.
Some procedures use hazardous voltages. To prevent
electrical shock, always set voltage source outputs to 0 V before
making or changing any interconnections.
Summary of Contents for 070-9731-05
Page 12: ...Getting Started ...
Page 13: ......
Page 24: ...Operating Basics ...
Page 25: ......
Page 59: ...General Purpose Application Examples 2 34 THS710A THS720A THS730A THS720P User Manual ...
Page 79: ...Power Measurement Application Examples 2 54 THS710A THS720A THS730A THS720P User Manual ...
Page 80: ...Reference ...
Page 81: ......
Page 83: ...Introduction to Reference 3 2 THS710A THS720A THS730A THS720P User Manual ...
Page 155: ...VERTICAL Controls 3 74 THS710A THS720A THS730A THS720P User Manual ...
Page 156: ...Appendices ...
Page 157: ......
Page 181: ...Appendix B Factory Setup B 4 THS710A THS720A THS730A THS720P User Manual ...
Page 187: ...Appendix C Accessories C 6 THS710A THS720A THS730A THS720P User Manual ...
Page 211: ...Appendix D Performance Verification D 24 THS710A THS720A THS730A THS720P User Manual ...
Page 213: ...Appendix E General Care and Cleaning E 2 THS710A THS720A THS730A THS720P User Manual ...
Page 214: ...Glossary and Index ...
Page 215: ......