Functional Overview
THS710A, THS720A, THS730A & THS720P User Manual
2–11
3. If necessary, adjust the probe for correct compensation.
P6117
P5102
4. Repeat these steps for the other probe and channel.
Compensating the Oscilloscope Signal Path
Signal path compensation optimizes the oscilloscope accuracy for
the current ambient temperature. For maximum accuracy, recompen-
sate the signal path if the ambient temperature changes by 5
°
C or
more.
1. Disconnect any probes or cables from the channel 1 and
channel 2 input BNC connectors.
2. Press UTILITY.
3. Select CAL from the System Menu.
4. Press Signal Path.
5. Press OK Compensate Signal Path. This procedure takes about
one minute to complete.
Summary of Contents for 070-9731-05
Page 12: ...Getting Started ...
Page 13: ......
Page 24: ...Operating Basics ...
Page 25: ......
Page 59: ...General Purpose Application Examples 2 34 THS710A THS720A THS730A THS720P User Manual ...
Page 79: ...Power Measurement Application Examples 2 54 THS710A THS720A THS730A THS720P User Manual ...
Page 80: ...Reference ...
Page 81: ......
Page 83: ...Introduction to Reference 3 2 THS710A THS720A THS730A THS720P User Manual ...
Page 155: ...VERTICAL Controls 3 74 THS710A THS720A THS730A THS720P User Manual ...
Page 156: ...Appendices ...
Page 157: ......
Page 181: ...Appendix B Factory Setup B 4 THS710A THS720A THS730A THS720P User Manual ...
Page 187: ...Appendix C Accessories C 6 THS710A THS720A THS730A THS720P User Manual ...
Page 211: ...Appendix D Performance Verification D 24 THS710A THS720A THS730A THS720P User Manual ...
Page 213: ...Appendix E General Care and Cleaning E 2 THS710A THS720A THS730A THS720P User Manual ...
Page 214: ...Glossary and Index ...
Page 215: ......