3. Set the TekScope instrument VOLTS/DIV to 100 mV/div.
4. Set the TekScope instrument SEC/DIV to 10
s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-
peak measurement is between 599 mV and 601 mV.
7. Set the leveled sine wave generator frequency to 200 MHz if you
are checking a THS730A, 100 MHz if you are checking a
THS720A or a THS720P, or 60 MHz if you are checking a
THS710A.
8. Press MAG.
9. Set the TekScope instrument SEC/DIV to 10 ns/div.
10. Check that the peak-to-peak measurement is
≥
425 mV.
11. Press MAG.
12. Proceed to the next test to check the channel 2 bandwidth.
Summary of Contents for 070-9731-05
Page 12: ...Getting Started ...
Page 13: ......
Page 24: ...Operating Basics ...
Page 25: ......
Page 59: ...General Purpose Application Examples 2 34 THS710A THS720A THS730A THS720P User Manual ...
Page 79: ...Power Measurement Application Examples 2 54 THS710A THS720A THS730A THS720P User Manual ...
Page 80: ...Reference ...
Page 81: ......
Page 83: ...Introduction to Reference 3 2 THS710A THS720A THS730A THS720P User Manual ...
Page 155: ...VERTICAL Controls 3 74 THS710A THS720A THS730A THS720P User Manual ...
Page 156: ...Appendices ...
Page 157: ......
Page 181: ...Appendix B Factory Setup B 4 THS710A THS720A THS730A THS720P User Manual ...
Page 187: ...Appendix C Accessories C 6 THS710A THS720A THS730A THS720P User Manual ...
Page 211: ...Appendix D Performance Verification D 24 THS710A THS720A THS730A THS720P User Manual ...
Page 213: ...Appendix E General Care and Cleaning E 2 THS710A THS720A THS730A THS720P User Manual ...
Page 214: ...Glossary and Index ...
Page 215: ......