General-Purpose Application Examples
THS710A, THS720A, THS730A & THS720P User Manual
2–29
Setup to Use External Trigger
SCOPE
TRIGGER
NU
Trigger Type
Edge
—
MENU
Trigger
Source
External
Connect the index pulse to the meter inputs, which now function as
the external trigger input. Adjust the TRIGGER LEVEL rocker for
stable triggering on the index pulse signal.
Going Further
You can use the external trigger input to trigger on a 50 Hz or 60 Hz
AC power line. This leaves the two channel inputs free to probe
other circuits that are synchronized to the AC power line. Set the
external trigger level to 0.2 V to trigger as close as possible to a zero
crossing.
Summary of Contents for 070-9731-05
Page 12: ...Getting Started ...
Page 13: ......
Page 24: ...Operating Basics ...
Page 25: ......
Page 59: ...General Purpose Application Examples 2 34 THS710A THS720A THS730A THS720P User Manual ...
Page 79: ...Power Measurement Application Examples 2 54 THS710A THS720A THS730A THS720P User Manual ...
Page 80: ...Reference ...
Page 81: ......
Page 83: ...Introduction to Reference 3 2 THS710A THS720A THS730A THS720P User Manual ...
Page 155: ...VERTICAL Controls 3 74 THS710A THS720A THS730A THS720P User Manual ...
Page 156: ...Appendices ...
Page 157: ......
Page 181: ...Appendix B Factory Setup B 4 THS710A THS720A THS730A THS720P User Manual ...
Page 187: ...Appendix C Accessories C 6 THS710A THS720A THS730A THS720P User Manual ...
Page 211: ...Appendix D Performance Verification D 24 THS710A THS720A THS730A THS720P User Manual ...
Page 213: ...Appendix E General Care and Cleaning E 2 THS710A THS720A THS730A THS720P User Manual ...
Page 214: ...Glossary and Index ...
Page 215: ......