Adjustments in the SCI Mode 
SCI OFF: At this adjustment a transfer of measured values from the E3907 is disabled. The E3907 can be 
controlled however with the E3907-commands.  
 
HAND: In this mode a measured value is issued via the serial interface when the Enter key is pressed at the E3907.  
 
AUTO: In this mode a measured value is issued via the serial interface in the adjusted delay  
 
TRIGGER: At a trigger event In this mode, a measured value is written in the logger. Since the trigger pulses may 
occur in 10ms raster and additional time-logg is not possible. The flank of the trigger pulse must be on HIGH for at 
least 4ms. Then it must be on LOW for at least 6ms. 
 
WINDOW: This mode reacts to increasing, respectively decreasing flanks. With a decreasing flank the logging of 
the measured values is started. Now the measured values are written in the data logger in 1ms raster. A decreasing 
flank ends the recording. 

 
4.6.4 Operation and function principle of the E3907-Data Logger  

 
The data logger can, if the E3907 is not in the measuring mode, be read by the menu option 2,7 LOGG - SENDING 
or by the command "A" via the interface. Outside of the measuring mode the data logger is deleted only by the 
menu option 2,7 LOGG - DELETION. 
 
If the E3907 is in the measuring mode, the data logger can be read with the command "A" and be deleted with the 
command "B". 
 
If the measuring mode is being switched on from the menu option 1MEASURING-sensor selection, 
the starting time of the measurement, the current sensor designation, the final value of the measuring range,  the 
adjusted measuring rate and the logger mode (e.g.: AUTO 1ms) are saved in the logger. 
 
NOTE: All previous measured values are deleted here! 
 
With the read-out of the data logger via the serial interface different adjustments are sent from the E3907.   
 

a) Company header 

 

b) Starting time of the measurement 

 

c) Sensor designation 

 

d) Final display  value 

 

e) Adjusted measuring rate 

 

f) Adjusted logger mode 

 

g) Thereafter the measured values 

 
In the HAND MODE the measured values are always logged with the time. The measured value is logged at  the 
data in the AUTOMODE. A time can be assigned to each measured value by the indicated starting time. 
Since in GRAPHS and in the WINDOW MODE trigger events smaller than 1sec can occur, an additional time log is 
not possible. 
 
Adjustments in the Logg Mode 
LOGG OUT:   Here the data logger is switched off. The logg mode in the measuring mode is switched to  "LOGG 

OUT" as soon as the entire data logger has been edited. 

 
HAND:  

In this mode a measured value is written into the data logger when the enter key was pressed at the 
E3907. By an additional logg of time, there is a time assignment for each measured value. 

 
AUTO :  

In this mode in the adjusted Delay a measured value is written into the data logger. By the stored 
starting time there is a time assignment for each measured value. 

 
GRAPH:  

In this mode at a trigger event a measured value is written into the logger. Since the trigger pulses 
occur in 10ms raster, an additional logg of the time is not possible. The flank of the trigger pulse must 
stand on HIGH for 4ms at least. Afterwards on LOW for at least 6ms. 

 

Summary of Contents for E3907

Page 1: ...Force Pressure Temperature Switch Service Operating manual E3907 Universal handheld measuring device with datalogger BD 984 i 02 2013...

Page 2: ...y 63073 Offenbach am Main Valid for E3907 handheld measuring device with datalogger Copyright 2006 2015 tecsis GmbH Offenbach Germany Reproduction prohibited Reproduction in whole or in part only with...

Page 3: ...interface 11 4 6 1 Transfer rate specification RS232 V 24 USB 11 4 6 2 Operation via the serial interface 11 4 6 3 Handling with serial interface USB 14 4 6 4 Operation and function principle of the E...

Page 4: ...Operating manual E3907 BE 984 i www tecsis de...

Page 5: ...arantee Errors and changes are reserved Intended Purpose Improper Usage A tecsis device is used for displaying processing and controlling or regulation of processes It shall not be used as the only to...

Page 6: ...take responsibility for that residual risks and to keep them as low as possible The relevant regulations and laws must be observed Residual risks or dangers associated with the measurement device shou...

Page 7: ...s The device may only be used by qualified personnel according to the technical data in connection with the special safety rules and regulations 1 8 Technical changes The product may not be changed co...

Page 8: ...An assignment of sensors to the corresponding parameter set is documented on the rear of the device 3 2 Power supply Mains operation Power supply 6V DC min 1 8A Fuse Self reset able fuse 1 5A Under Ov...

Page 9: ...evice from moisture condensation rain snow Protect the device from direct sunlight Protect the device from dust and dirt Protect the device from high or excessive ambient temperature Protect the devic...

Page 10: ...nu ESC Switch off the device Exit one step back in the menu Scroll up Scroll down USB USB Socket for USB Connector Type B RS232 RS232 Socket for 3 5mm jack 3pol front GND middle RxD outside TxD 6V DC...

Page 11: ...Operating manual E3907 www tecsis de BE 984 i 7...

Page 12: ...Operating manual E3907 8 BE 984 i www tecsis de 4 2 Switching on the device Press the ENTER button to switch on the device 4 3 Menu description 4 3 1 Brief handling overview E3907...

Page 13: ...w many measurements an average value is accumulated Applicable at e g vibrations control oscillations 2 5 PASS Password query After entering the password 9373 it is possible to enter menu 3 CALIB Ther...

Page 14: ...appear on the LCD Press ENTER again and the time can be adjusted With PLUS and MINUS the hours can be adjusted After ENTER the minutes can be adjusted with PLUS and MINUS With ENTER the selected time...

Page 15: ...uring every keystroke on ENTER while measuring For data logging adjust operating mode in the logger mode e g choose AUTO and interval time e g 10 sec and change to measuring mode A measured value with...

Page 16: ...Out datalogger B 0x42 Delete datalogger C 0x43 Read Out current sensor parameter D 0x44 Read Out status E 0x45 Read Out complete status g 0x67 Change protocol setup k 0x6B ENTER l 0x6C PLUS m 0x6D MI...

Page 17: ...mA and 100 load mA Dec point 0xXXXX XXXX 000 _5000___ 001 _5 000__ 010 _50 00__ 011 _500 0__ 100 5 000___ 0 load 2Byte HEX value MSB LSB 100 load 2Byte HEX value MSB LSB 4 6 2 3 Read out status Status...

Page 18: ...ol status 1Byte 0xXXXX XXXX binary coded 1 no closing character is sent 1 send CR LF 1 send CR 1 send LF 4 6 2 12 Calculation of the Checksum CS and the Weighted Checksum wCS The calculation takes pla...

Page 19: ...starting time of the measurement the current sensor designation the final value of the measuring range the adjusted measuring rate and the logger mode e g AUTO 1ms are saved in the logger NOTE All pr...

Page 20: ...ress key or and choose between 0 Load or Sensitivity 0 Load Relieve sensor from load or Sensitivity Enter the value fort the sensitivity in V Press Key or and choose between 100 Load or Sensitivity 10...

Page 21: ...0mA short circuit protected Bridge resistance min 350 Ohm only full bride 4 or 6 conductor Sensitivity up to 3 3mV V other values on request 100 control via E3907 applied via control input 4 8 2 Conne...

Page 22: ...measures before installation Attention The information for the voltage supply in section 5 2 must be respected Cable connections Attention Never connect voltage supply on non used pins 5 3 Normal ope...

Page 23: ...k the plug connections Repair Notation The products contain no parts that need to be or that can be serviced by the user himself Repairs may only be carried out by the tecsis GmbH If it can be assumed...

Page 24: ...tecsis GmbH Carl Legien Stra e 40 44 63073 Offenbach am Main Germany Phone 49 69 5806 0 Fax 49 69 5806 7788 E Mail force tecsis de Internet www tecsis de...

Reviews: