Operating manual
E3907
BE 984 i
9
4.3.2 Menu description
1 MEASURE Measuring mode
SENSOR__
Sensor selection for measuring mode, sensor 0-9
Here the sensor parameter set for the measurement is being selected.
The sensor parameter set must match the connected sensor.
2 SYSTEM
All system parameters are being stored in this menu column.
2.1 LANG
Language adjustment
The menu language can be chosen in German, English, French or Spanish
2.2 INFO
Information query was configured factory-sided an is not variable
VERS NO.
Software version
2.3 SCI
Interface configuration
BAUD
Baud rate adjustment must be conform with the receiver (pc or printer).
SCI MODE
Interface configuration only refers to the measuring mode, not the logg mode!
OFF
Interface off
HAND
At key stroke ENTER the measured value is issued.
AUTO¹
Interval time, adjustable from 10ms to 1h (at high measuring rate only possible with highest baud
rate)
TRIGGER
At impulse on trigger input the measured value is issued.
INTERFAC
Auswahl der zu verwendenden Schnittstelle
USB
USB interface is activated
RS232
RS232 interface is activated
2.4 RATE
Measuring rate and average value
RATE
Measuring rate adjustment can be chosen between 1/s and 1000/s. At fast courses (screw joint,
insert press...) always choose a fast measuring rate, e.g. 1000/s, at very slow courses a small rate,
e.g. 1/s.
AVERAGE
Forming an average value, the numbers indicate by how many measurements an average value is
accumulated. Applicable at e.g. vibrations, control oscillations....
2.5 PASS
Password query
After entering the password 9373 it is possible to enter menu 3 CALIB . There the sensor
parameters can be modified.
2.6 LCD
LCD-contrast adjustment
If the LCD becomes unreadable by external influences e.g. solar radiation, heat or cold, the LCD-
contrast can be corrected here by using the or key
2.7 LOGG
Data logger adjustments
LOGGMODE Data logger configuration.
OFF
Off-switch for logger operation, by this the measuring mode is possible only.
HAND
At each key stroke on ENTER a logg value is being taken over.
AUTO
Adjustment of the measuring intervals.
DIAGRAM
With each increasing flank of the trigger signal a logg value is being taken over.
WINDOW
The window operation is being started by an increasing trigger flank and can only be finished by a
decreasing flank.
LOGGSEND
Logger values are issued through the interface, press ENTER twice.
LOGGDEL
Logger values are deleted, beforehand there will be a safety query whether the buffer should be
deleted.
2.8 DATE
Date and time adjustment
DATE
Date
TIME
Time
3 CALIB
Calibration menu
this is active only, if the password has been entered correctly
in 2.5PASS. For the calibration the sensor needs to be connected to the sensor socket.
¹
If the SCI → AUTO interval
is shorter than the set measuring rate, then the interface displays the same output until
a new value is recorded.
Sensor__
Sensor number or name of the sensor which is to be calibrated.
POINTDIS
Measuring range with decimal indication of the sensor, e.g. 100,0; 200,0; 63,00;...